![Rohde & Schwarz CMU-K20 Operating Manual Download Page 335](http://html.mh-extra.com/html/rohde-and-schwarz/cmu-k20/cmu-k20_operating-manual_1477804335.webp)
CMU-K20/K21/K22/K23/K24
Measurement Groups (Non Signalling only)
1115.6088.12 6.31
E-12
CALCulate[:SCALar]POWer[:NORMal][:GMSK]:MPR:LIMit:MATChing?
Tolerance Error
Returned values
Value range
Def. value
Def. unit
FW vers.
AvgBurstPwCurr,
PeakBurstPwCurr,
BurstMatching,
NMAU | NMAL | INV | OK
NMAU | NMAL | INV | OK
INV | MATC | NMAT | OUT | NTR | NRAM |
OFLW | UFLW | NTSC | OFF
INV
INV
INV
–
–
–
V2.00
PhErrPeakCurrent,
PhErrPeakAverage,
PhErrPeakMaxMin,
PhErrRMSCurrent,
PhErrRMSAverage,
PhErrRMSMaxMin,
FreqErrCurrent,
FreqErrAverage,
FreqErrMaxMin,
AvgBurstPwAvg
For all measured values:
NMAU | NMAL | INV | OK
INV
INV
INV
INV
INV
INV
INV
INV
INV
INV
–
–
–
–
–
–
–
–
–
–
Description of command
This command is always a query. It indicates whether and in which way the tolerances for the scalar results (see
command above) in the
Power
and the
modulation
measurement have been exceeded.
The following messages may be returned for the values A
vgBurstPower
and
PeakBurstPower
and for all results
of the modulation measurement:
NMAU
Tolerance value underflow
not matching, underflow
NMAL
Tolerance value exceeded
not matching, overflow
INV
Measurement
invalid
invalid
OK
Tolerance value matched
The following messages may be returned for the value
BurstMatching
:
INV
invalid
MATC
matching
NMAT
not matching
OUT
out of range
NTR
no trigger
NRAM
not ramping (burst not found)
OFLW
overflow
UFLW
underflow
NTSC
no training sequence code
OFF
off
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...