![Rohde & Schwarz CMU-K20 Operating Manual Download Page 339](http://html.mh-extra.com/html/rohde-and-schwarz/cmu-k20/cmu-k20_operating-manual_1477804339.webp)
CMU-K20/21/22/23/24 Common:
WPOWer
1115.6088.12 6.35
E-12
STATus:OPERation:SYMBolic:ENABle <Event>{,<Event>}
Symbolic status evaluation
Parameter list
Parameter description
Def. Value
1
Default Unit
FW vers.
<Event>{,<Event>} |
NONE
List of symbols for events to be reported
No event reported
NONE –
V3.05
Command description
This command enables event reporting for one or several events in the current
GSMxxx-MS Non Signalling
function group, i.e. it sets the corresponding bits in the
STATus:OPERation:CMU:SUM<nr>:CMU<nr_event>:ENABle
register
(<nr> = 1 | 2, <nr_event>
denotes the current function group) and in all sum registers up to the status byte. The events and the
corresponding symbols for the function group are listed in Chapter 5 (see section
Status Registers).
The symbols
may be entered in arbitrary order.
STATus:OPERation:SYMBolic[:EVENt]?
Symbolic status evaluation
Response
Parameter description
Def. Value
2
Default Unit
FW vers.
NONE |
<Event>{,<Event>}
No event in the
RF
function group
List of reported events
NONE –
V3.05
Command description
This command is always a query. It lists the events reported in the current
GSMxxx-MS Non Signalling
function
group and deletes these events in the
STATus:OPERation:CMU:SUM<nr>:CMU<nr_event>:EVENt
register
as well as in all sum registers.
WPOWer
The subsystem
WPOWer
measures the power of the signal transmitted by the mobile phone using a
wideband filter. It corresponds to the softkey
Wideband Power
in the
Connect. Control
menu.
INITiate:WPOWer
Start new measure ment
⇒
RUN
ABORt:WPOWer
Abort measurement and switch off
⇒
OFF
STOP:WPOWer
Stop measurement
⇒
STOP
CONTinue:WPOWer
Next measurement step (only
counting mode
)
⇒
RUN
Description of command
FW vers.
These commands have no query form. They start or stop the measurement, setting it to the status
given in the top right column.
1.15
3
1
The default values quoted in this command are achieved after a
STATus:PRESet
command.
*RST
does not overwrite the entries in the status registers; see section
Reset
Values of the Status Reporting Systems
in chapter 5.
2
The default values quoted in this command are achieved after a
*CLS
command.
*RST
does not overwrite the entries in the status registers; see section
Reset Values of
the Status Reporting Systems
in chapter 5.
3
For firmware versions <V2.15, the keyword
WPOWer
is replaced by
SPOWer
in all commands.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...