![Rohde & Schwarz CMU-K20 Operating Manual Download Page 314](http://html.mh-extra.com/html/rohde-and-schwarz/cmu-k20/cmu-k20_operating-manual_1477804314.webp)
Config. File Management – System MMEMory
CMU-K20/K21/K22/K23/K24
1115.6088.12 6.10
E-12
Config. File Management – System MMEMory
The
MMEMory
system provides mass storage capabilities for the CMU. The functionality of this system
is included in the
Data
menu; see CMU200/300 operating manual.
The mass storage of the CMU may be internal or external. The internal mass storage device is a sec-
tion on the internal hard disk that is reserved for mass storage (directory c:\temp). The external mass
storage device is either a floppy disk or a PCMCIA memory card, depending on the instrument configu-
ration. The
<msus>
(mass storage unit specifier) parameter in the
MMEMory
commands denotes the
root directory of the
INTernal
or
EXTernal
mass storage device.
The <FileName> parameter is a string. The contents of the string may contain characters for specifying
subdirectories, e.g. "\TEMP\TRASH\test.txt" for the file named
test.txt
in the
TEMP\TRASH
subdirectory
of the root directory or "TEMP\TRASH\test.txt" for the file named
test.txt
in the
TEMP\TRASH
subdirec-
tory of the current directory, to be queried with the base system command
MMEMory:DIRectory
[:CURRent]?
. The file name itself may contain the period as a separator for extensions.
MMEMory:SAVE:CURRent
<FileName> [,<msus>]
Save configurations in current function group and test mode
Parameters
Parameter description
Def. value
Def. unit
FW vers.
"<FileName>",
INTernal | EXTernal
Name of the config. file to be created
Storage device of the config. file
–
INTernal
–
–
V3.10
Command description
This command saves the configuration of the current function group and test mode to a configuration file. A "?" in
the specified file name will be replaced by current numbers that are automatically incremented, starting with zero.
The auto-increment function overwrites an existing file with a "9" in its file name. For instrument settings that may
be different in manual and remote control (e.g. the repetition mode for many measurements) the manual setting
is saved. The command is available in all function groups. This command is CMU-specific.
MMEMory:RECall:CURRent
<FileName> [,<msus>]
Recall configurations in current function group and test mode
Parameters
Parameter description
Def. value
Def. unit
FW vers.
"<FileName>",
INTernal | EXTernal
Name of the config. file to be recalled
Storage device of the config. file
–
INTernal
–
–
V3.10
Command description
This command recalls the configuration of the current function group and test mode from a configuration file. The
command is available in all function groups. This command is CMU-specific.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...