CMU-K20/21/22/23/24 Connection
Control (Signalling only)
1115.6088.12 6.195
E-12
CONFigure:NETWork:SYSTem:CACCess
<Mode>
Cell Access
Mode>
Description of parameters
Def. value
Def. unit
FW vers.
BARRed |
NBARred
Radio cell disabled for all mobiles
Radio cell accessible
NBARred – V1.15
Description of command
Sig. State
This command enables or disables the radio cell for mobiles.
SOFF,
SON
Q: all
CONFigure:NETWork:SYSTem:BINDicator
<Band>
Band Indicator
<Blocks>
Description of parameters
Def. value
Def. unit
FW vers.
G18 | G19
GSM band that the mobile can use
see below
–
V3.10
Description of command
Sig. State
This command sets the band indicator of the MS under test. The default value is G18 if the current
function group is GSM400/900 or GSM1800 and G19 if the current function group is GSM850 or
GSM1900.
≠
CEST,
Q: all
CONFigure:NETWork:SYSTem:BSAGblksres
<Blocks>
Number of Reserved Blocks
<Blocks>
Description of parameters
Def. value
Def. unit
FW vers.
0 to 7
Number of reserved blocks in the BCCH for access
channel
0 – V1.15
Description of command
Sig. State
This command determines the number of data blocks to be reserved for the granted access
(access
grant channel = AGC)
within the BCCH
(basic services access grant blocks reserved)
.
≠
CEST,
Q: all
CONFigure:NETWork:SYSTem:BSPamfrms
<Frames>
Paging Request Interval
<Frames>
Description of parameters
Def. value
Def. unit
FW vers.
2 to 9
Interval between two paging requests
2
V1.15
Description of command
Sig. State
This command defines the interval between two paging requests in a multiframe in frames
(basic
services paging blocks available per multiframe)
.
≠
CEST,
Q: all
CONFigure:NETWork:SYSTem:BSPReorganis
<Mode>
Paging
Reorganisation
<Mode>
Description of parameters
Def. value
Def. unit
FW vers.
ON | OFF
Paging reorganization on/off
OFF
–
V2.00
Description of command
Sig. State
This command switches the paging reorganization parameter on and off. In the
OFF
state, the mobile
is prevented from switching to the idle mode.
≠
CEST,
Q: all
CONFigure:NETWork:SYSTem:PLUPdate
<Value>
TS1232, Periodic Location Update
<Value>
Description of parameters
Def. value
Def. unit
FW vers.
OFF |
0 to 255
No periodic location update performed
Value of T1232 timer
OFF deci-
hours
V3.10
Description of command
Sig. State
This command sets the value of the timer T3212 of the periodic location updating procedure. The unit
decihours corresponds to 6 minutes or 360 seconds.
≠
CEST,
Q: all
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...