CMU-K20/K21/K22/K23/K24
Non Signalling: Power Measurements
1115.6088.12 4.29
E-11
Remote control
CONFigure:POWer[:NORMal][:GMSK]:CONTrol:DEFault ON | OFF
etc.
Repetition
The
Repetition
parameter
defines how often the measurement is repeated:
Single Shot
Single-shot measurement: the measurement is stopped after a
statistics cycle (or after a stop condition is met, see below). A
stopped measurement is indicated by the status display
HLT
in
the
Power
softkey.
Unless otherwise stated, a statistics cycle corresponds to the
number of bursts/evaluation periods set under
Statistic Count
.
Continuous
Continuous measurement: The CMU continues the
measurement until it is terminated explicitly (or until the stop
condition for the measurement is met, see below). The
measurement results are valid after one statistics cycle;
however, the measurement is continued, and the output is
continuously updated. An ongoing measurement is indicated by
the status display
RUN
in the softkey
Power
.
Single shot should be selected if only a single measurement result is required
under fixed conditions. The continuous measurement is suitable for monitoring the
evolution of a measured quantity in time, for example for adjustments.
Note:
In remote mode, the counting measurement (counting mode) is
available as a further measurement mode with a defined number of
measurement cycles to be performed, see chapter 6 of this manual.
Remote control
CONFigure:POWer[:NORMal][:GMSK]:CONTrol:REPetition
CONTinuous | SINGleshot | 1 ... 10000,<StopCondition>,
<Stepmode>
etc.
Stop Condition
The
Stop Condition
field
defines a stop condition for the measurement:
NONE
Continue measurement irrespective of the results of the limit
check
On Limit Failure
Stop measurement as soon as the limit check fails (one of the
tolerances is exceeded)
Remote control
CONFigure:POWer[:NORMal][:GMSK]:CONTrol:REPetition
<Repetition>,SONerror | NONE, <Stepmode> etc.
Display Mode
The
Display Mode
field
defines which of the four measured and calculated
measurement curves is displayed. The measurement curves differ in the way the
burst power
p(t)
at a fixed point in time
t
is calculated if the measurement extends
over several bursts:
Current
Measured value for current burst
Minimum
Minimum over all measured bursts
Maximum
Maximum over all measured bursts
Average
Average value over a number of bursts
The number of bursts for calculation of the statistical values
Minimum, Maximum
and
Average
– and thus the result – depends on the repetition mode set. In detail,
this implies:
Single shot
Display of minimum, maximum and average value from the
performed statistics cycle.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...