![Rohde & Schwarz CMU-K20 Operating Manual Download Page 508](http://html.mh-extra.com/html/rohde-and-schwarz/cmu-k20/cmu-k20_operating-manual_1477804508.webp)
Connection Control (Signalling only) CMU-K20/21/22/23/24
1115.6088.12 6.204
E-12
OUTPut[:STATe]
<State>
RF Output
<State>
Description of parameters
Def. value
Def. unit
FW vers.
RF1
RF2
RF3
Connector RF1 used as output
Connector RF2 used as output
Connector RF3 OUT used as output
RF2 – V1.15
Description of command
Sig. State
This command determines the connector to be used for RF output signals. The bidirectional
connectors RF1 and RF2 can be used as input and output connectors in the same measurement (see
INPut[:STATe]
).
Only one input and one output may be active at the same time, a new RF output setting overwrites
the previous one.
all
[SENSe:]CORRection:LOSS:INPut<nr>[:MAGNitude]
<Attenuation >
Ext. Att. Input
SOURce:CORRection:LOSS:INPut<nr>[:MAGNitude]
<Attenuation >
<Attenuation>
Description of parameters
Def. value
Def. unit
FW vers.
–50 dB to +50 dB
–90 dB to +90 dB
Ext. attenuation at RF <nr>, where <nr> = 1,2
Ext. attenuation at RF 4 IN
0 dB
V1.15
Description of command
Sig. State
This command assigns an external attenuation value to one of the inputs defined before (see
command
INPut:STATe)
.
all
[SENSe:]CORRection:LOSS:OUTPut<nr>[:MAGNitude]
<Attenuation>
Ext. Att. Output
SOURce:CORRection:LOSS:OUTPut<nr>[:MAGNitude]
<Attenuation>
<Attenuation>
Description of parameters
Def. value
Def. unit
FW vers.
–50 dB to +50 dB
–90 dB to +90 dB
Ext. attenuation at RF <nr>, where <nr> = 1,2
Ext. attenuation at RF3 OUT
0 dB
V1.15
Description of command
Sig. State
This command assigns an external attenuation value to one of the outputs defined before (see
command
OUTPut:STATe
).
all
ROUTe:SPENcoder[:INPut]
<Source>
Speech Encoder
<Source>
Description of parameters
Def. value
Def. unit
FW vers.
HANDset |
GENerator
Handset is used as source
AF generator is used as source
HAND
– V2.00
Description of command
Sig. State
This command determines the input source that feeds the CMU speech encoder (option CMU-B52).
all
ROUTe:SPDecoder[:OUTPut]
<Destination>
Speech Decoder
<Destination>
Description of parameters
Def. value
Def. unit
FW vers.
HANDset |
ANALyzer |
ANA2 |
ABOTh
Speech decoder output routed to the handset
Speech dec. output routed to primary AF analyzer
Speech dec. output routed to secondary AF analyzer
Speech dec. output routed to both AF analyzers
HAND –
V2.00
Description of command
Sig. State
This command routes the CMU speech decoder output (option CMU-B52). The ANA2 and ABOTh
settings are provided in firmware versions
≥
3.05.
all
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...