Non Signalling Mode
CMU-K20/-K21/-K22/-K23/-K24
1115.6088.12 2.12
E-10
⇓
Step 6
ESCAPE
!
Press the ESCAPE key to close the
Power Configuration
menu and return to
the main menu.
⇓
The trace is now continuously measured
and updated in the display. With the dis-
play mode
Maximum,
which is indicated in
the upper right corner of the diagram, trace
values will be replaced only if a current
measured value at a particular test point
exceeds all values measured before at the
same test point.
⇓
Step 7
!
Press the
Menus
softkey to display the
measurement groups available in the
hotkey bar.
!
Press the
Modulation
hotkey to open the
Modulation
menu.
⇓
The
Modulation
menu displays the results
of the phase and frequency error meas-
urement.
"
The trace represents the phase error of the
current burst as a function of time.
#
Below, a table displays the extreme value
of the phase error and the RMS phase
error, the origin offset, the I/Q imbalance,
and the frequency error.
$
The detected training sequence (TSC),
average power of the current burst and the
statistic count are shown in addition.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...