CMU-K20/K21/K22/K23/K24
Non Signalling: Power Measurements
1115.6088.12 4.11
E-11
P/t Normal
8PSK
The
P/t Normal 8PSK
hotkey selects the power versus time measurement for
8PSK modulated normal burst signals (see explanation of GSM burst structure at
the beginning of section
Limit lines (Power Configuration – Limit Lines)
on page
4.32
).
Remote control:
The
P/t Normal 8PSK
application is selected by the keywords
[:NORMal]:EPSK
in
the 3
rd
and 4
th
level of the POWer commands, e.g.
CONFigure:POWer
[:NORMal]:EPSK...
P/t
Multislot
The
P/t Multislot
hotkey selects the power versus time measurement for multislot
configurations (see explanation of GSM burst structure at the beginning of section
Limit lines (Power Configuration – Limit Lines)
on page
4.32
).
Remote control:
The
P/t Multislot
application is selected by the 3
rd
level keyword :
MSLot
in the
POWer commands, e.g.
CONFigure:POWer:MSLot...
P/Frame
The
P/Frame
hotkey selects the power versus frame measurement. In this
application, the average burst power in a particular timeslot is measured over a
range of consecutive TDMA frames and displayed in tabular form.
Remote control:
The
P/Frame
application is selected by the keyword
:FRAMe
in the 3
rd
level of the
POWer
commands, e.g.
CONFigure:POWer:FRAMe...
P/Slot
Graph
The
P/Slot Graph
hotkey selects the power versus slot measurement with
graphical display. In this application, the average burst power in all eight timeslots
of a TDMA frame is measured and displayed in a bar graph.
Remote control:
The
P/Slot Graph
application is selected by the keyword
:SLOT
in the 3
rd
level of
the
POWer
commands, e.g.
CONFigure:POWer:SLOT...
P/Slot
Table
The
P/Slot Table
hotkey selects the power versus slot measurement with tabular
display. In this application, the average burst power in all eight timeslots of several
consecutive TDMA frames is measured and displayed in a table.
Remote control:
The
P/Slot Table
application is selected by the keyword
:XSLot
in the 3
rd
level of
the
POWer
commands, e.g.
CONFigure:POWer:XSLot...
Some of the following test settings depend on the selected application.
c) P/t Normal GMSK
All softkeys and hotkeys in the
P/t Normal GMSK
application are shown in
Fig. 4-2
on page
4.9
.
P/t Norm.
GMSK
The
P/t Norm. GMSK
measurement control softkey controls the
P/t Norm. GMSK
measurement; see detailed explanation in section
Measurement Control
on p.
4.10
.
Besides, the measurement control
softkey provides hotkeys to define the scope of
the measurement. All these settings are described in more detail in section
Measurement Control (Power Configuration – Control)
on page
4.28
ff.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...