![Rohde & Schwarz CMU-K20 Operating Manual Download Page 127](http://html.mh-extra.com/html/rohde-and-schwarz/cmu-k20/cmu-k20_operating-manual_1477804127.webp)
CMU-K20/K21/K22/K23/K24
Non Signalling: Modulation Measurements
1115.6088.12 4.49
E-11
Parameter line
The first parameter line contains the following settings:
Max. Level
Maximum input level set as in the
Analyzer
tab of the
Connection Control
menu
(see section
Table-Oriented
Version
on p.
4.73
ff.)
Attenuation
Setting for the external attenuation of the input level
(Normal,
Low Noise, Low Distortion)
Freq. Offset
Frequency offset compared to the nominal channel frequency
Chan./Trig. Slot Offs.
RF channel and trigger slot offset (see
Trig. Slot Offset
hotkey on p.
4.18
)
Remote control
The settings are retrieved using the query corresponding to the setting command
(setting command with appended question mark).
Output fields
Below the diagram, the following results are displayed:
TSC detected
Detected training sequence of the current burst received from the
mobile station
(GSM 0 to 7
or
Dummy
or
"---"),
see
Analyzer
tab,
section
Table-Oriented Version
on p.
4.73
ff.
95
th
percentile
Limit values below which 95% of the measured
Error Vector
Magnitudes, Magnitude Errors,
and
Phase Errors
in the current
burst are located. Owing to this definition, the 95
th
percentile of a
measured quantity has the same unit as the quantity itself.
Output table
The output table contains the following scalar values:
Avg. Burst Power
Average power of current burst (irrespective of the display mode
selected and of the current trace),
Statistic Count
Number of sweeps per statistics cycle. The colored bar indicates
the relative measurement progress in the statistics cycle,
Bursts out of
Percentage of bursts that violate the tolerance limits.
Tolerance
The following scalar values are calculated for the current burst first. From the
current results the average referenced to a statistics cycle (
Average,
see averaging
rule in Chapter 3, section
General Settings)
and the extreme value over all bursts
measured so far
(Max/Min)
is calculated:
Err. Vect. Magn. (Peak)
Maximum EVM (application
EVM 8PSK
only)
Err. Vect. Magn. (RMS)
Effective EVM (RMS-averaged over the burst)
Magn. Error (Peak)
Maximum magnitude error (application
Magn. Err. 8PSK
only)
Magn. Error (RMS)
Effective magnitude error (RMS-averaged over the burst)
Phase Error (Peak)
Maximum phase error (application
Phase Err. 8PSK
only)
Phase Error (RMS)
Effective phase error (RMS-averaged over the burst)
Origin Offset
Origin offset in the I/Q constellation diagram
Frequency Error
Difference between measured and expected signal
frequency
Peak and RMS values are specific to the current application
(Phase Error,
Magnitude Error
or
Error Vector Magnitude)
. For an explanation of all quantities
measured refer to the beginning of this section on page
4.43
.
Limit Check
A red output field and an arrow pointing upwards or downwards indicates that the
measurement result exceeds the upper or lower limit set in the
Limits
tab of the
TX
Tests
configuration menu, see p.
4.52
.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...