Preparing a GSM Mobile Phone Test
CMU-K20/-K21/-K22/-K23/-K24
1115.6088.12 2.2
E-10
Preparing a GSM Mobile Phone Test
This chapter describes how to use the CMU for GSM mobile phone tests. As a prerequisite for starting
the session, the instrument must be correctly set up and connected to the AC power supply as de-
scribed in chapter 1 of the operating manual for the CMU basic unit. Furthermore, the GSM software
must be properly installed following the instructions given in chapter 1 of the present manual.
o
I
Step 1
!
Switch on the CMU using the mains
switch at the rear.
"
!
Check the operating mode of the instru-
ment at the
ON/STANDBY
key on the
front panel.
#
⇓
0
1
2
3
4
5
6
7
8
9
.
-
ON / OFF
ENTER
1100.0008.02
VARIATION
DATA
FUNCTION
SYSTEM
CONTROL
DATA
CTRL
MENU
SELECT
HELP
SETUP
mark
symb
E
F
B
M/µ
µV
W
C
dBµV
k/m
G/n
A
mV
D
1
dB
dBm
UNIT...
EXP/CMP
CONT/HALT
RF4 IN
UNIVERSAL RADIO COMMUNICATION TESTER CMU 200
.
ESCAPE
*
abc
def
ghi
jkl
mno
pqr
stu
vwx
yz
_ µ
AF IN
AF OUT
VOL
DEL
AUTO
INFO
RESET
*
#
Ω
AUX 1
AUX 2
SPEECH
AUX 3
DATA 2
13 dBmMAX
2 W MAX
50 W MAX
13 dBmMAX
RF 3 OUT
RF 2
RF 1
DATA 1
INS
CLR
2
DEF
3
GHI
1
ABC
5
6
4
8
ÜVW
7
STU
.
-
0
9
XYZ
S
C
RCL
M
Step 2
!
Connect the bi-directional RF connector
RF 2 of the CMU to the antenna connec-
tor of the mobile phone.
$
!
Supply the mobile phone with the correct
operating voltage (battery or power sup-
ply).
%
⇓
Step 3
!
Switch on the CMU by means of the
ON/STANDBY
key on the front panel.
⇓
The startup menu is displayed while the
CMU performs a power-up test.
&
After a few seconds the CMU displays the
last menu used in the previous session.
!
Press the
RESET
key to open the
Reset
popup menu.
!
Select
Reset
and press the
ENTER
key.
!
In the popup window opened, select
Yes
to confirm the instrument reset.
The CMU indicates that it performs a gen-
eral reset of all device settings and is then
ready to carry out the following steps. The
Reset
popup menu is closed automatically.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...