Common: MODulation:EVMagnitude CMU-K20/21/22/23/24
1115.6088.12 6.90
E-12
READ:ARRay:MODulation:EVMagnitude:EPSK[:CURRent]?
Phase Error in Burst
READ:ARRay:MODulation:EVMagnitude:EPSK:AVERage?
READ:ARRay:MODulation:EVMagnitude:EPSK:MMAXimum?
Start single shot measurement and return results
⇒
RUN
FETCh:ARRay:MODulation:EVMagnitude:EPSK[:CURRent]?
FETCh:ARRay:MODulation:EVMagnitude:EPSK:AVERage?
FETCh:ARRay:MODulation:EVMagnitude:EPSK:MMAXimum?
Read measurement results (unsynchronized)
⇒
RUN
SAMPle:ARRay:MODulation:EVMagnitude:EPSK[:CURRent]?
SAMPle:ARRay:MODulation:EVMagnitude:EPSK:AVERage?
SAMPle:ARRay:MODulation:EVMagnitude:EPSK:MMAXimum?
Read measurement results (synchronized)
⇒
RUN
Returned values
Description of parameters
Def. value
Def. unit
FW vers.
0.0 % to+ 100.0 %,
... ,
0.0 % to+ 100.0 %
1
st
value for error vector magnitude
...
xth value for error vector magnitude
NAN
...
NAN
%
...
%
V2.15
Description of command
These commands are always queries. They return the error vector magnitude vs. time at fixed, equidistant test
points. The number of measured values is 588, corresponding to a time range of 0 bit to 146 ¾ bit.
The calculation of
current, average, and mmax
(Min./Max.) results is explained in chapter 3 (see
display mode
).
READ:SUBarrays:MODulation:EVMagnitude:EPSK[:CURRent]?
Subarray Results
READ:SUBarrays:MODulation:EVMagnitude:EPSK:AVERage?
READ:SUBarrays:MODulation:EVMagnitude:EPSK:MMAXimum?
Start single shot measurement and return results
⇒
RUN
FETCh:SUBarrays:MODulation:EVMagnitude:EPSK[:CURRent]?
FETCh:SUBarrays:MODulation:EVMagnitude:EPSK:AVERage?
FETCh:SUBarrays:MODulation:EVMagnitude:EPSK:MMAXimum?
Read meas. results (unsynchronized)
⇒
RUN
SAMPle:SUBarrays:MODulation:EVMagnitude:EPSK[:CURRent]?
SAMPle:SUBarrays:MODulation:EVMagnitude:EPSK:AVERage?
SAMPle:SUBarrays:MODulation:EVMagnitude:EPSK:MMAXimum?
Read results (synchronized)
⇒
RUN
Ret. values per subrange
Description of parameters
Def. value
Def. unit
FW vers.
0.0 % to+ 100.0 %,
... ,
0.0 % to+ 100.0 %
1
st
value for error vector magnitude
...
xth value for error vector magnitude
NAN
...
NAN
%
...
%
V2.15
Description of command
These commands are always queries. They measure and return the error vector magnitude versus time in the
subranges defined by means of the
CONFigure:SUBarrays:MODulation:EVMagnitude:EPSK
command.
In the default setting of the configuration command the
READ:SUBarrays..., FETCh:SUBarrays..., and
SAMPle:SUBarrays...
command group is equivalent to the
READ:ARRay..., FETCh:ARRay..., and
SAMPle:ARRay...
command group described above.
The
CONFigure:SUBarrays:MODulation:EVMagnitude:EPSK
command defines a maximum of 32
subranges. If one of the statistical modes
(ARIThmetical, MINimum, MAXimum)
is set, only one value is
returned per subrange.
The calculation of
current, average, minimum,
and
maximum
results is explained in chapter 3 (see
display
mode
).
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...