Non Signalling: Power Measurements
CMU-K20/K21/K22/K23/K24
1115.6088.12 4.28
E-11
Measurement Configurations (Power Configuration)
The popup menu
Power Configuration
contains three tabs to determine the parameters controlling the
power measurement including the error tolerances.
The popup menu
Power Configuration
is activated by pressing the measurement control softkey at the
top right in the graphical measurement menu
Power
twice. It is possible to change between the tabs by
pressing the associated hotkeys.
Measurement Control (Power Configuration – Control)
The
Control
tab controls the power measurement by determining
•
The
Repetition
mode
•
The
Stop Condition
for the measurement (for burst power vs. time measurements only)
•
The type of measurement curve displayed
(Display Mode,
for burst power vs. time measurements
only)
•
The number of bursts/evaluation periods forming a statistics cycle
(Statistic Count,
for burst power
vs. time measurements only)
•
The
measurement
Filter
for
P/t Normal GMSK, P/t Normal 8PSK
and
P/t Multislot
measurements
•
The averaging rule to obtain the reference power
(Ref. Power Mode
, for
8PSK
measurements only)
•
The
Timing Offset
, the expected modulation
(Modulation View)
,
the number of slots measured
(Slot
Count)
and the
Info Box
in the
P/t Multislot
application
•
The number of timeslots measured
(Slot Count)
in the
P/Slot Table
application
•
The number of frames measured
(Frame Count)
in the
P/Frame
application
Besides, it configures the graphical diagrams by adding or removing the
Grid
.
Fig. 4-10 Power Configuration – Control
The settings can be defined independently for the different applications of the
Power
measurement
group. The following settings are available in several applications:
Default Settings
The
Default All Settings
switch assigns default values to all settings in the
Control
tab (the default values are quoted in the command description in chapter 6 of this
manual). In addition, default switches for the individual modulation schemes are
provided.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...