Common: POWer:MSLot
CMU-K20/21/22/23/24
1115.6088.12 6.58
E-12
Subsystem SUBarrays:POWer:MSLot
The subsystem
SUBarrays:POWer:MSLot
defines the measurement range and the type of output
values.
CONFigure:SUBarrays:POWer:MSLot
Definition of Subarrays
<Mode>,<Start>,<Samples>{,<Start>,<Samples>}
<Mode>
Description of parameters
Def. value
Def. unit
ALL |
ARIThmetical |
MINimum |
MAXimum |
IVAL,
Return all measurement values
Return arithm. mean value in every range
Return minimum value in every range
Return maximum value in every range
Return single interpolated value at <Start>
ALL –
<Start>
Description of parameters
Def. value
Def. unit
–180 symbols to
+520 symbols
Start time in current range, relative to symbol 0 of
the meas. slot
–165 symb.
<Samples>
Description of parameters
Def. value
Def. unit
FW vers.
1 to 2613
Number of samples in current range, depending on
SCOunt (
see commands
READ:ARRay:POWer:MSLot
...)
2613 – V3.05
Description of command
This command configures the
READ:SUBarrays:POWer..., FETCh:SUBarrays:POWer...,
and
SAMPle:SUBarrays:POWer
commands. It restricts the measurement to up to 32 subranges where either all
measurement results (the number of which is given by the second numerical parameter) or a single statistical
value is returned. The subranges are defined by the start time and the number of test points which are located on
a fixed, equidistant grid with a step width of ¼ symbols. If <Start> does not coincide with a test point then the
range will start at the next test point that is larger than <Start>.
For <Mode> = IVAL, the <Samples> parameter is ignored and the CMU returns a single measurement value at
the abscissa value <Start>. If <Start> is located beween two test points with valid results then the result is
calculated from the results at these two adjacent test points by linear interpolation.
The subranges may overlap but must be within the total range of the
POWer
measurement. Test points outside
this range are not measured (result
NAN)
and do not enter into the ARIThmetical, MINimum and MAXimum
values.
By default, only one range corresponding to the total measurement range is used and all measurement values
are returned.
Measured Values
The subsystem
POWer:MSLot...
contains the commands to measure the normal burst power, compare
it with the tolerances and retrieve the results. The subsystem corresponds to the graphical
measurement menu
Power.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...