CMU-K20/K21/K22/K23/K24
Non Signalling: Connection Control
1115.6088.12 4.75
E-11
Analyzer Level –
RF Attenuation
The
RF Attenuation
parameter
defines how the RF analyzer of the CMU is tuned to
meet the requirements of the current measurement type. In general, a compromise
between the acceptable noise level in the displayed result and the contribution of
internally generated distortion must be reached.
Normal
Mixer level in normal range,
Low noise
Mixer level enhanced by +10 dB (full dynamic range of CMU,
therefore recommended for
Power
and
Spectrum
measurements),
Low distortion
Mixer level reduced by –10 dB (high intermodulation spacing,
therefore recommended for
Modulation
measurements).
The
Attenuation
setting permits the CMU to be adapted to the requirements of the
measurement. The advantages and disadvantages of the settings
Low noise
and
Low distortion
are listed in the following table.
Advantages
Disadvantages
Low noise
Low noise
high dynamic range
No RF overdrive reserve
Risk of intermodulation
Low distortion
High intermodulation spacing
Lower dynamic range
Remote control
[SENSe:]LEVel:ATTenuation NORMal | LNOise | LDIStortion
Meas. Control –
Multi Slot –
Trigger Slot
Offset
Trig. Slot Offset
defines a delay time (integer number of GSM timeslots) between
the trigger time and the timeslot that is measured in all
Multislot
configurations; see
Fig. 4-3
on p.
4.18
.
Remote control
CONFigure:RFANalyzer:MCONTrol:TSOFfset 0 to 7
Analyzer
Settings –
Template PCL
Template PCL
sets a power control level to correct the limit lines. See
Template
PCL
softkey on p.
4.13
.
Remote control
CONFigure:RFANalyzer:TPCL <PCL>
Generator Settings (Connection Control – Generator)
The
Generator
tab controls and configures the RF generator. The CMU provides a softkey-oriented
version of the
Generator
tab and a table-oriented version with extended functionality. The
Generator
hotkey toggles between the two versions if it is pressed repeatedly.
Softkey-Oriented Version
The softkey-oriented version of the
Generator
tab provides the following RF generator settings:
•
Generator control and level in the used and unused timeslots (measurement control softkey
RF
Generator
)
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...