CMU-K20/21/22/23/24 Signalling:
POWer[:NORMal]
1115.6088.12 6.127
E-12
CONFigure:POWer[:NORMal][:GMSK]:LIMit:LINE:UPPer<AreaNr>:ALL:DYNamic:ENABle
CONFigure:POWer[:NORMal]:EPSK:LIMit:LINE:UPPer<AreaNr>:ALL:DYNamic:ENABle
<Enable>
Dynamic Correction on/off
<Enable>
Description of parameters
Def. value Def. unit
FW vers.
ON | OFF
Switch dynamic correction on or off
ON
–
V2.00
Description of command
This command switches the dynamic correction of the upper limit area <nr> for all ten PCL ranges on or off.
CONFigure:POWer[:NORMal][:GMSK]:LIMit:LINE:UPPer:ALL:DYNamic:ENABle
<Enable>
CONFigure:POWer[:NORMal]:EPSK:LIMit:LINE:UPPer:ALL:DYNamic:ENABle
<Enable>
Dynamic Correction on/off
<Enable>
Description of parameters
Def. value Def. unit
FW vers.
ON | OFF
Switch dyn. correction for the whole template on or off
ON
–
VV2.00
Description of command
This command switches the dynamic correction of the upper limit line in all areas and for all PCL ranges on or
off. The query returns 160 Boolean values corresponding to the limit check in PCL ranges 1 to 10 (inner loop) in
each of the areas 1 to 16 (outer loop).
CONFigure:POWer[:NORMal][:GMSK]:LIMit:LINE:LOWer<AreaNr>:DYNamic<RangeNr>
CONFigure:POWer[:NORMal]:EPSK:LIMit:LINE:LOWer<AreaNr>:DYNamic<RangeNr>
<fromTPCL>, <toTPCL>, <Correction>, <Enable>
Dynamic Correction
<fromPCL>
Description of parameters
Def. value
Def. unit
0 to 31 |
MAX |
OFF
First PCL for which area <AreaNr> is changed
OFF
PCL
<toPCL>
Description of parameters
Def. value
Def. unit
0 to 31
|
MAX |
OFF
Last PCL for which area <AreaNr> is changed.
OFF
PCL
<Correction>
Description of parameters
Def. value
Def. unit
-10 dB to
+10 dB
|
OFF
Correction value (relative) for the limit line in area
<AreaNr> applied for all PCLs between <fromPCL> and
<toPCL> (including)
OFF dB
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON | OFF
Enable or disable dynamic correction in the current limit
line area and PCL range
OFF –
VV2.00
Description of command
These command activates and defines dynamic correction of the lower limit line of area <AreaNr> depending on
the PCL range <RangeNr>.
MAX
denotes the maximum output power (smallest PCL) of the mobile phone under
test.
By default, the dynamic limit line correction is disabled in all ranges and areas.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...