![Rohde & Schwarz CMU-K20 Operating Manual Download Page 123](http://html.mh-extra.com/html/rohde-and-schwarz/cmu-k20/cmu-k20_operating-manual_1477804123.webp)
CMU-K20/K21/K22/K23/K24
Non Signalling: Modulation Measurements
1115.6088.12 4.45
E-11
The I/Q imbalance is measured for GMSK-modulated signals only.
Fig. 4-19 Measurement menu Modulation – Overview 8PSK
Test Settings
The
Analyzer Level, Analyzer Settings, Generator,
and
Menus
test settings are identical with those in
the
Power
menu
(see section
Test settings
on page
4.9
)
.
The
Overview 8PSK
measurement control
softkey (which changes to
EVM 8PSK, Magn. Error 8PSK,
or
Phase Error 8PSK
if the corresponding
application is selected) is analogous to the
Ext. Phase Err. GMSK
softkey described in section
Test
Settings
on page
4.39
. With 8PSK modulation, the
Application
softkey provides the following
applications:
Application
The
Application
softkey selects the measurement application and the modulation
scheme. Several applications of the
Modulation
menu are related to 8PSK
modulated signals. The GMSK application is described in section
Test Settings
on
p.
4.39
f.
Overview
8PSK
The
Overview 8PSK
hotkey selects all scalar modulation results to be displayed.
For an explanation of the measured quantities see section
Measurement Menu
(Modulation – 8PSK)
on page
4.42
.
Remote control
No explicit switchover command. All
Overview 8PSK
measurements are identified
by the 3
rd
/4
th
level keywords
...OVERview:EPSK...
EVM
8PSK
The
EVM 8PSK
hotkey selects the magnitude of the error vector to be displayed.
The error vector connects the measured signal from the mobile station and the
ideal signal vector at the symbol points, see explanation in section
Measurement
Menu (Modulation – 8PSK)
on page
4.42
. The diagram shows the relative
magnitude (in percent), i.e. the ratio of the magnitude of the error vector to the
magnitude of the ideal signal vector.
Remote control
No explicit switchover command. All
EVM 8PSK
measurements are identified by
the 3
rd
/4
th
level keywords
...EVMagnitude:EPSK...
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...