Non Signalling: Analyzer/Generator Menu
CMU-K20/K21/K22/K23/K24
1115.6088.12 4.4
E-11
Frequency
The
Frequency
hotkey opens the input window
Frequency
.
Input windows indicate the current parameter value (in this
case: the current RF input frequency) or a list of the
possible settings. Parameters are changed by
•
Overwriting/incrementing numerical values (for
numerical parameters)
•
Selecting from the list of parameters (for select
parameters)
Measurement Control
Each
Analyzer/Generator
application is controlled by means of the measurement control softkey below
the
Connect. Control
softkey and the associated hotkeys.
P/t Norm.
GMSK
The
P/t Norm. GMSK
softkey (which changes to
Ext. Phase Err. GMSK
or
Overview 8PSK,
depending on the application selected) controls the measurement
application and indicates its status
(RUN | HLT | OFF)
. This status can be changed
after softkey selection (pressing once) by means of the
ON/OFF
key or the
CONT/HALT
key. The status can be set independently for all three applications.
The applications
P/t Norm. GMSK
and
Ext. Phase Err. GMSK
can be run in
parallel, so the results for both applications are displayed simultaneously.
Switchover between these two applications does not change the course of the
measurement.
The GMSK applications and the
Overview 8PSK
suspend each other. The selected
measurement status of each application is stored and will be put into effect as soon
as the application is activated. In particular, an application in the status
RUN
is
restarted each time it is activated.
Remote control
INITiate:POWer[:NORMal][:GMSK]
etc.
FETCh:POWer[:NORMal][:GMSK]:STATus?
INITiate:MODulation:XPERror[:GMSK]
etc.
FETCh
:
MODulation:XPERror[:GMSK]:STATus?
INITiate:MODulation:OVERview:EPSK
etc.
FETCh:MODulation:OVERview:EPSK:STATus?
Measurement
configuration
The configuration menus for all
Power
and
Modulation
measurements are directly
accessible from the
Analyzer/Generator
menu:
•
Pressing
the
P/t Norm. GMSK
softkey twice opens the popup menu
Power
Configuration
(see page
4.28
ff.).
•
Pressing
the
Ext. Phase Err. GMSK
or the
Overview 8PSK
softkey twice opens
the popup menu
Modulation Configuration
(see page
4.50
ff.).
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...