CMU-K20/K21/K22/K23/K24
Signalling: Receiver Quality Measurements
1115.6088.12 4.139
E-12
(automatic gain control). The
AGC Holdoff Time
is also applied
if the RF level changes during the
Receiver Quality
measurement.
Sync Holdoff Time
Delay time after which the signalling unit of the CMU starts
synchronizing to the bit stream. This holdoff time is not applied
in the case of RF level changes during the
Receiver Quality
measurement.
Both holdoff times depend on the performance of the mobile station. A reduction of
the holdoff times towards their mobile-dependent lower limits can accelerate the
measurement.
Remote control
CONFigure:RXQuality:CONTrol:HTIMe <AGCTime>,<SynchTime>
Confidence
Settings
The
Confidence Settings
section sets the parameters for statistical BER tests. For
background information and application examples refer to section
Statistical BER
Tests
on p.
4.125
ff. The settings are available for the
BER
application and valid in
the measurement modes
RBER/FER, BER
and
Burst by Burst.
Note:
In the measurement modes
RBER/FER
and
BER,
the Class II BER
provides the pass/fail criteria. In the measurement modes
Burt by
Burst
and
BER/DBLER
, the
BER
calculated from all bits is used. No
statistical results are provided for AMR full rate tests and for
measurement in GPRS test mode A.
Statistical testing is activated by setting
Stop Condition = Confidence Level;
see
below.
The results
(Early Fail, Early Pass etc.)
are displayed in the measurement
menu. The following confidence settings are provided:
Confidence Fail
Confidence level for early fail decisions: After the
Min. Test Time
,
the confidence test is stopped with an
”Early Fail”
message as
soon as the probability of the DUT to have a BER above the
specified upper limit (see section
Upper Limits for Bit Error Rate
(Receiver Quality Configuration – Limits)
on p.
4.143
ff.) is larger
than the confidence fail value.
Confidence Pass
Confidence level for early pass decisions: After the
Min. Test
Time
, the confidence test is stopped with an
”Early Pass”
message
as soon as the probability of the DUT to have a BER
below the specified upper limit
times the
Bad DUT
factor
of 1.5
is larger than the confidence pass value.
Note:
The confidence levels are the complement of the
probabilities of
wrong decision F
used in the test specification:
<Confidence Level> = 1 – F.
Bad DUT
Fixed bad DUT factor of 1.5; see section
Statistical BER Tests
on p.
4.125
ff.
Result Window
BER range factor for statistical dual-limit BER tests. A factor of
n
% means that the BER must be within the range
[<BER Limit> (1 – n %), <BER Limit> (1 + n %)]. The
OFF
setting
corresponds to a single-limit BER test.
Min. Test Time
Minimum test time before a check of the early pass and early fail
limits can stop the measurement. Minimum test times are
necessary in particular if the test conditions introduce some
fluctuations that disturb the statistical independence of the single
bit error events and must be averaged out (e.g. multipath fading,
birth/death propagation, moving propagation).
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...