![Rohde & Schwarz CMU-K20 Operating Manual Download Page 417](http://html.mh-extra.com/html/rohde-and-schwarz/cmu-k20/cmu-k20_operating-manual_1477804417.webp)
CMU-K20/21/22/23/24 Common: SPECtrum:SWITching
1115.6088.12 6.113
E-12
CONFigure:SPECtrum:SWITching[:GMSK]:CONTrol <
Mode>,<Statistics
>
CONFigure:SPECtrum:SWITching:EPSK:CONTrol <
Mode>,<Statistics
>
Scope of Measurement
<Mode>
Description of parameters
Def. value
Def. unit
SCALar |
ARRay
Only scalar measured values
Scalar measured values and arrays
ARRay –
<Statistics>
Description of parameters
Def. value
Def. unit
FW vers.
1 to 1000 |
NONE
Number of bursts per statistics cycle
Statistics off (equivalent to 1)
10 – V1.20
Description of command
This command restricts the type of measured values and determines the number of bursts within a statistics cy-
cle.
CONFigure:SPECtrum:SWITching[:GMSK]:CONTrol:REPetition
CONFigure:SPECtrum:SWITching:EPSK:CONTrol:REPetition
Test Cycles
<
Repetition
>
,<StopCondition>,
<Stepmode>
<Repetition>
Description of parameters
Def. value
Def. unit
FW vers.
CONTinuous |
SINGleshot |
1 to 10000
Continuous measurement (until
STOP
or
ABORT)
Single shot measurement (until
Status = RDY
)
Multiple measurement
(
counting
, until
Status = STEP | RDY)
SING –
<StopCondition>
Description of parameters
Def. value
Def. unit
SONerror |
NONE
Stop measurement in case of error
(stop on error)
Continue measurement even in case of error
NONE –
<Stepmode>
Description of parameters
Def. value
Def. unit
FW vers.
STEP |
NONE
Interrupt measurement after each statistics cycle
Continue measurement according to its rep. mode
NONE –
V1.20
Description of command
This command defines the number of test cycles, the stepping mode and, if required, a stop condition for the
measurement.
Note:
In the case of READ commands (
READ:
...), the <Repetition> parameter has no effect; the measure-
ment is always stopped after a single shot
DISPlay:SPECtrum:SWITching[:GMSK]:CONTrol:GRID
<Enable>
DISPlay:SPECtrum:SWITching:EPSK:CONTrol:GRID
<Enable>
Grid
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON |
OFF
Switch on grid lines
Switch off grid lines
ON – V1.20
Description of command
This command switches the grid lines in the test diagrams on or off.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...