CMU-K20/21/22/23/24 Common:
MODulation:PERRor
1115.6088.12 6.95
E-12
Measured Values – Subsystem MODulation[:PERRor]:EPSK
The subsystem
MODulation[:PERRor]:EPSK
measures and returns the modulation parameters and
compares them with the tolerance values. The subsystem corresponds to the various output elements
in the measurement menu
MODulation,
application
Phase Error 8PSK.
Scalar
Results:
READ[:SCALar]:MODulation[:PERRor]:EPSK?
Start single shot meas. and return results
FETCh[:SCALar]:MODulation[:PERRor]:EPSK?
Read out meas. results (unsynchronized)
SAMPle[:SCALar]:MODulation[:PERRor]:EPSK?
Read out measurement results (synchronized)
Returned values
Value range
Def. value
Def. unit
FW vers.
95thPercentilePhase Error
PhErrorPeak (x3),
PhErrorRMS (x3),
OriginOffset (x3),
FrequencyError (x3),
AvgBurstPowerCurr,
BurstsOutOfTol
0.0 % to 100.0 %
0.0 % to 100.0 %
0.0 % to 100.0 %
–100.0 dB to +100.0 dB
–1000.0 Hz to +1000.0 Hz
–137 dBm to +53 dBm
0.0 % to 100.0 %
NAN
NAN
NAN
NAN
NAN
NAN
NAN
%
%
%
dB
Hz
dBm
%
V2.15
Description of command
These commands are always queries. They start a modulation measurement and output all scalar measurement
results (see chapter 4), either for the whole burst or for the 1
st
ten valid symbols in the burst. The calculation of
results in an
average
or
peak
measurement is described in chapter 3 (see
calculation of statistical quantities
).
The symbol (x3) behind a value indicates that the list contains three results corresponding to the
Current
, the
Average,
and the
MMax
value.
CALCulate:MODulation[:PERRor]:EPSK:LIMit:MATChing?
Bursts out of Tolerance
Returned values
Value range
Def. value
Def. unit
FW vers.
95thPercentilePhError
PhErrorPeak (x3),
PhErrorRMS (x3),
OriginOffset (x3),
FrequencyError(x3)
For all measured values:
NMAU | NMAL | INV | OK
INV
INV
INV
INV
INV
–
–
–
–
–
V2.15
Description of command
This command is always a query. It indicates whether and in which way the error limits for the scalar measured
values (see above command) have been exceeded. The symbol (x3) behind a value indicates that the list con-
tains three results corresponding to the
Current
, the
Average,
and the
MMax
value. The limits are defined with
the
CONFigure:MODulation:OEMP
... commands.
The following messages may be output for all measured values:
NMAU
Underflow of tolerance value
not matching, underflow
NMAL
Tolerance value exceeded
not matching, overflow
INV
Measurement invalid
invalid
OK
all tolerances matched
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...