CMU-K20/K21/K22/K23/K24
Connection Control (Non Signalling only)
1115.6088.12 6.7
E-12
Subsystem for Input and Output (Connectors, External Attenuation)
The subsystem for input and output configures the input and output connectors. The subsystem corre-
sponds to the tab
RF
in the popup menu
Connect. Control
.
INPut[:STATe]
<State>
RF Input
<State>
Description of parameters
Def. value
Def. unit
FW vers.
RF1 |
RF2 |
RF4
Connector RF 1 used as input
Connector RF 2 used as input
Connector RF 4 IN used as input
RF2 –
V1.15
Description of command
This command determines the connector to be used for RF input signals. The bidirectional connectors RF 1 and
RF 2 can be used both as input and output connectors in the same measurement (see
OUTPut[:STATe]
).
Only one input and one output may be active at the same time, a new RF input setting supersedes the previous
one.
OUTPut[:STATe]
<State>
RF Output
<State>
Description of parameters
Def. value
Def. unit
FW vers.
RF1 |
RF2 |
RF3
Connector RF 1 used as output
Connector RF 2 used as output
Connector RF 3 OUT used as output
RF2 –
V1.15
Description of command
This command determines the connector to be used for RF output signals. The bidirectional connectors RF 1 and
RF 2 can be used as input and output connectors in the same measurement (see
INPut[:STATe]
).
Only one input and one output may be active at the same time, a new RF output setting supersedes the previous
one.
[SENSe:]CORRection:LOSS:INPut<nr>[:MAGNitude]
<Attenuation>
SOURce:CORRection:LOSS:INPut<nr>[:MAGNitude]
<Attenuation>
Ext.
Att.
Input
<Attenuation>
Description of parameters
Def. value
Def. unit
FW vers.
–50 dB to +50 dB
–90 dB to +90 dB
Value for ext. attenuation at input<nr> where <nr> = 1,
2
Value for ext. attenuation at RF4 IN (<nr> = 4)
0.0
0.0
dB
dB
V1.15
Description of command
This command assigns an external attenuation value to the inputs of the instrument
(RF 1, RF 2, RF4 IN)
.
[SENSe:]CORRection:LOSS:OUTPut<nr>[:MAGNitude]
<Attenuation>
SOURce:CORRection:LOSS:OUTPut<nr>[:MAGNitude]
<Attenuation>
Ext.
Att.
Output
<Attenuation>
Description of parameters
Def. value
Def. unit
FW vers.
–50 dB to +50 dB
–90 dB to +90 dB
Value for ext. attenuation at output<nr> where <nr> =
1, 2
Value for ext. attenuation at RF3 OUT (<nr> = 3)
0.0
0.0
dB
dB
V1.15
Description of command
This command assigns an external attenuation value to the outputs of the instrument
(RF 1, RF 2, RF3 OUT)
.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...