Common: NPOWer
CMU-K20/21/22/23/24
1115.6088.12 6.38
E-12
NPOWer
The subsystem
NPOWer
measures the power of the signal transmitted by the mobile phone using the
RF analyzer configuration of the
POWer
measurement. Compared to
WPOWer
, the
NPOWer
measurement uses a narrow-band (500 kHz Gauss) filter.
The narrow-band
NPOWer
measurement yields the average, maximum and minimum burst power of
the current burst. In addition to these
Current
values the minimum and maximum power in the entire
measurement and the average of the average current values, referenced to a statistics cycle, is
calculated. The entire measurement curves (arrays) are not available, and no limit check is performed.
NPOWer
is a quick and precise alternative to the
WPOWer
or
POWer
measurements if only scalar
results are needed.
Note:
A Free Run trigger (
TRIGger[:SEQuence]:SOURce FRUN
) should be avoided because
it delays the
NPOWer
measurement.
INITiate:NPOWer
Start new measurement
⇒
RUN
ABORt:NPOWer
Abort measurement and switch off
⇒
OFF
STOP:NPOWer
Stop measurement
⇒
STOP
CONTinue:NPOWer
Next measurement step (only
counting mode
)
⇒
RUN
Description of command
FW vers.
These commands have no query form. They start or stop the measurement, setting it to the status
given in the top right column.
V3.05
CONFigure:NPOWer:EREPorting <
Mode
>
Event
Reporting
<Mode>
Description of parameters
Def. value
Def. unit
FW vers.
SRQ |
SOPC |
SRSQ |
OFF
Service request
Single operation complete
SRQ and SOPC
No reporting
OFF –
V3.05
Description of command
This command defines the events generated when the measurement is terminated or stopped
(event reporting,
see chapter 5 of CMU200 operating manual).
FETCh:NPOWer:STATus?
Measurement
Return
Description of parameters
Def. value
Def. unit
FW vers.
OFF |
RUN |
STOP |
ERR |
STEP |
RDY,
1 to 10000 |
NONE
1 to 1000 |
NONE
Measurement in the
OFF
state (
*RST
or
ABORt
)
Running (after
INITiate, CONTinue
or
READ
)
Stopped (
STOP
)
OFF
(could not be started)
Stepping mode
(<stepmode>=STEP)
Stopped according to repetition mode and stop condition
Counter for current statistics cycle
No counting mode set
Counter for current evaluation period within a cycle
Statistic count set to off
OFF
NONE
NONE
–
–
–
V3.05
Description of command
This command is always a query. It returns the status of the measurement (see chapters 3 and 5 of CMU manual
operating manual).
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...