![Rohde & Schwarz CMU-K20 Operating Manual Download Page 376](http://html.mh-extra.com/html/rohde-and-schwarz/cmu-k20/cmu-k20_operating-manual_1477804376.webp)
Common: MODulation:XPERror
CMU-K20/21/22/23/24
1115.6088.12 6.72
E-12
MODulation:XPERror
The subsystem
MODulation:XPERror[:GMSK]
measures the modulation parameters (frequency and
phase errors) in GMSK modulation including the I/Q Imbalance and the Origin Offset in the I/Q
constellation diagram. The subsystem corresponds to the measurement menu
Modulation,
application
Ext. Phase Err. GMSK,
and the associated popup menu
Modulation Configuration.
If no results for the I/Q Imbalance and the Origin Offset are needed, it is recommended to use the faster
MODulation[:PERRor]
measurement; see p.
6.64
ff.
Control of Measurement – Subsystem MODulation:XPERror[:GMSK]
The subsystem
MODulation:XPERror[:GMSK]
controls the modulation measurement. It corresponds to
the softkey
Ext. Phase Err. GMSK
in the measurement menu
Modulation
.
INITiate:MODulation:XPERror[:GMSK]
Start new measurement
⇒
RUN
ABORt:MODulation:XPERror[:GMSK]
Abort running measurement and switch off
⇒
OFF
STOP:MODulation:XPERror[:GMSK]
Stop measurement after current stat. cycle
⇒
STOP
CONTinue:MODulation:XPERror[:GMSK]
Next measurement step (only
stepping mode
)
⇒
RUN
Description of command
FW vers.
These commands have no query form. They start and stop the modulation measurement, setting it
to the status indicated in the top right column.
V2.15
CONFigure:MODulation:XPERror[:GMSK]:EREPorting <
Mode
>
Event
Reporting
<Mode>
Description of parameters
Def. value
Def. unit
FW vers.
SRQ |
SOPC |
SRSQ |
OFF
Service request
Single operation complete
SRQ and SOPC
No reporting
OFF –
V2.15
Description of command
This command defines the events generated when the measurement is terminated or stopped
(event reporting,
see chapter 5 of CMU manual).
FETCh:MODulation:XPERror[:GMSK]:STATus?
Measurement Status
Ret. values
Description of parameters
Def. value
Def. unit
FW vers.
OFF |
RUN |
STOP |
ERR |
STEP |
RDY,
1 to 10000 |
NONE,
1 to 1000 |
NONE
Measurement in the
OFF
state (
*RST
or
ABORt
)
Running (after
INITiate, CONTinue
or
READ
)
Stopped (
STOP
)
OFF
(could not be started)
Stepping mode
(<stepmode>=STEP)
Stopped according to repetition mode and stop condition
Counter for current statistics cycle
No counting mode set
Counter for current evaluation period within a cycle
Statistic count set to off
OFF
NONE
NONE
–
–
–
V2.15
Description of command
This command is always a query. It returns the status of the measurement (see chapters 3 and 5 of CMU
manual).
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...