Common: WPOWer
CMU-K20/21/22/23/24
1115.6088.12 6.36
E-12
CONFigure:WPOWer:EREPorting <
Mode
>
Event
Reporting
<Mode>
Description of parameters
Def. value
Def. unit
FW vers.
SRQ |
SOPC |
SRSQ |
OFF
Service request
Single operation complete
SRQ and SOPC
No reporting
OFF –
1.17
Description of command
This command defines the events generated when the measurement is terminated or stopped
(event reporting,
see chapter 5 of CMU200 operating manual).
FETCh:WPOWer:STATus?
Measurement
Return
Description of parameters
Def. value
Def. unit
FW vers.
OFF |
RUN |
STOP |
ERR |
STEP |
RDY,
1 ... 10000 |
NONE
Measurement in the
OFF
state (
*RST
or
ABORt
)
Running (after
INITiate, CONTinue
or
READ
)
Stopped (
STOP
)
OFF
(could not be started)
Stepping mode
(<stepmode>=STEP)
Stopped according to repetition mode and stop condition
Counter for current statistics cycle
No counting mode set
OFF
NONE
–
–
1.15
Description of command
This command is always a query. It returns the status of the measurement (see chapters 3 and 5 of CMU
operating manual).
CONFigure:WPOWer:CONTrol:REPetition <
Repetition
>
,<StopCond>,<Stepmode>
Test cycles
<Repetition>
Description of parameters
Def. value
Def. unit
CONTinuous |
SINGleshot |
1 ... 10000
Continuous measurement (until
STOP
or
ABORT)
Single shot measurement (until
Status = RDY
)
Multiple measurement
(
counting
, until
Status = STEP | RDY)
SING
–
<StopCond>
Description of parameters
Def. value
Def. unit
SONerror |
NONE
Start measurement in case of error
(stop on error)
Continue measurement even in case of error
NONE –
<Stepmode>
Description of parameters
Def. value
Def. unit
FW vers.
STEP |
NONE
Interrupt measurement after each statistics cycle
Continue measurement according to its rep. mode
NONE –
1.15
Description of command
This command determines the number of statistics cycles, the stop condition and the stepping mode for the
measurement.
Note:
In the case of READ commands (
READ:
...), the <Repetition> parameter has no effect; the
measurement is always stopped after a single shot.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...