CMU-K20/K21/K22/K23/K24
Measurement Groups (Non Signalling only)
1115.6088.12 6.27
E-12
Subsystem POWer:MPR:CONTrol
The subsystem
POWer:MPR:CONTrol
defines the repetition mode, statistic count, and stop condition of
the measurement.
CONFigure:POWer[:NORMal][:GMSK]:MPR:CONTrol <
Mode>, <Statistics
>
Scope of Measurement
<Mode>
Desciption of parameters
Def. value
Def. unit
SCALar |
ARRay
Scalar values only (incl. ramp matching)
Scalar measured values and arrays
ARRay –
<Statistics>
Description of parameters
Def. value
Def. unit
FW vers.
1 to 1000 |
NONE
Number of bursts per statistics cycle
Statistics off (equivalent to 1)
100 – V2.00
Description of command
This command specifies the type of measured values and defines the number of bursts forming a statistics cycle.
CONFigure:POWer[:NORMal][:GMSK]:MPR:CONTrol:REPetition
<
Repetition
>
,<StopCond>,<Stepmode>
Test Cycles
<Repetition>
Description of parameters
Def. value
Def. unit
CONTinuous |
SINGleshot |
1 to 10000
Continuous measurement (until
STOP
or
ABORT)
Single shot measurement (until
Status = RDY
)
Multiple measurement
(
counting
, until
Status = STEP | RDY)
SING –
<StopCond>
Description of parameters
Def. value
Def. unit
SONerror |
NONE
Stop measurement in case of error
(stop on error)
Continue measurement even in case of error
NONE –
<Stepmode>
Description of parameters
Def. value
Def. unit
FW vers.
STEP |
NONE
Interrupt measurement after each statistics cycle
Continue measurement according to its rep. mode
NONE –
V2.00
Description of command
This command determines the number of statistics cycles, the stop condition and the stepping mode for the
measurement.
Note:
In the case of READ commands (
READ:
...), the <Repetition> parameter has no effect; the measure-
ment is always stopped after a single shot.
Test Configuration
The commands of the following subsystems configure the combined power and modulation measure-
ment. Note that configurations such as limit lines must be defined separately in the
POWer
and
MODu-
lation
systems.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...