CMU-K20/21/22/23/24 Signalling: Receiver Quality
1115.6088.12 6.161
E-12
Returned values for BDBL
Value range
Def. value
Def. unit
FW vers.
ProgressTime,
BER,
USFBLER
DBLER
CRCErrors,
Status
Statistical Result*
)
0.0% to 100.0%
0.000% to 100.000%
0.000% to 100.000%
0.000% to 100.000%
0 to 50000
INV | PASS | FAIL | TLOW | IMP
RUN | EFAI | EPAS | FAIL | PASS | THIG |
TLOW
NAN
NAN
NAN
NAN
NAN
INV
INV
%
%
%
%
–
–
–
V2.00
V3.40
Description of command
These commands are always queries. They start a bit-error-rate test in the single shot repetition mode and output
the measurement results (see also detailed explanation of measured values in chapter 4 and table in section
BER Tests of PDTCHs
). The results depend on the measurement mode set via the
CONFigure:RXQuality:BER<nr>:CONTrol
command (RFER, BER, BBB, DBLer). They are
ProgressTime Relative
progress of the measurement
ClassIIBits
(Residual) bit error rate for class II bits
ClassIbBits
(Residual) bit error rate for class Ib bits
FER
Frame erasure rate
BER
Bit error rate (no distinction between bit classes)
DBLER
Data block error rate
USFBLER
USF block error rate, available for packet-switched channels only (CMU-K42)
CRCErrors
Cyclic redundancy check (CRC) errors
Status Measurement
status
The following messages can be output for the measurement status:
INV Measurement
invalid
invalid
PASS
All tolerances matched
passed
FAIL
Not all tolerances matched
failed
IMP Measurement
impossible, therefore invalid
impossible
*) The statistical result is available only if statistical BER testing is activated via
CONFigure:RXQuality:BER<nr>:CONTrol:REPetition
(see p.
6.156
) following messages can be output
for the measurement status:
INV Measurement
invalid
invalid
PASS
All tolerances matched
passed
FAIL
Not all tolerances matched
failed
IMP Measurement
impossible, therefore invalid
impossible
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...