CMU-K20/21/22/23/24 Signalling:
POWer[:NORMal]
1115.6088.12 6.121
E-12
CONFigure:POWer[:NORMal][:GMSK]:EREPorting <
Mode
>
Event
Reporting
CONFigure:POWer[:NORMal]:EPSK:EREPorting <
Mode
>
<Mode>
Description of parameters
Def. value
Def. unit
FW vers.
SRQ |
SOPC |
SRSQ |
OFF
Service request
Single operation complete
SRQ and SOPC
No reporting
OFF –
V1.15
Description of command
This command defines the events generated when the measurement is terminated or stopped
(event reporting,
see chapter 5 of CMU manual).
FETCh:POWer[:NORMal][:GMSK]:STATus?
Measurement Status
FETCh:POWer[:NORMal]:EPSK:STATus?
Returned values
Description of parameters
Def. value
Def. unit
FW vers.
OFF |
RUN |
STOP |
ERR |
STEP |
RDY,
1 to 10000 |
NONE,
1 to 1000 |
NONE
Measurement in the
OFF
state (
*RST
or
ABORt
)
Running (after
INITiate, CONTinue
or
READ
)
Stopped (
STOP
)
OFF
(could not be started)
Stepping mode
(<stepmode>=STEP)
Stopped according to repetition mode and stop condition
Counter for current statistics cycle
No counting mode set
Counter for current evaluation period within a cycle
Statistic count set to off
OFF
NONE
NONE
–
–
–
V1.15
Description of command
This command is always a query. It returns the status of the measurement (see chapters 3 and 5 of CMU
manual).
CONFigure:POWer[:NORMal][:GMSK]:TOFFset
<Offset>
Bit Offset
CONFigure:POWer[:NORMal]:EPSK:TOFFset
<Offset>
<Offset>
Description of parameters
Def. value
Def. unit
FW vers.
–4.00 to +4.00
Number of bits
0
bit
V2.15
Description of command
This command defines an offset time in ¼ bit units by which the burst is shifted relative to the time axis and the
tolerance template.
CONFigure:POWer[:NORMal][:GMSK]:FILTer
<Filter>
Filter
CONFigure:POWer[:NORMal]:EPSK:FILTer
<Filter>
<Mode>
Description of parameters
Def. value
Def. unit
FW vers.
G500 |
B600
500 kHz Gaussian filter
600 kHz bandpass filter
G500 for GMSK modulation
B600 for 8PSK modulation
– V3.05
Description of command
This command selects the measurement filter for the
P/t
measurement. The default filter setting differs for the
two modulation schemes.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...