![Rohde & Schwarz CMU-K20 Operating Manual Download Page 426](http://html.mh-extra.com/html/rohde-and-schwarz/cmu-k20/cmu-k20_operating-manual_1477804426.webp)
Signalling: POWer[:NORMal] CMU-K20/21/22/23/24
1115.6088.12 6.122
E-12
Test Configuration
The commands of the following subsystems configure the power measurement. They correspond to the
sections in the
Power Configuration
popup menu that are related to the normal burst power
measurement. For a detailed explanation of the power tolerance template defined in the GSM standard
see chapter 4.
Subsystem POWer[:NORMal]:CONTrol
The subsystem
POWer[:NORMal]:CONTrol
defines the repetition mode, statistic count, and stop
condition of the measurement. These settings are provided in the
Control
tab of the popup menu
Power
Configuration.
CONFigure:POWer[:NORMal][:GMSK]:CONTrol <
Mode>, <Statistics
>
Scope of Measurement
CONFigure:POWer[:NORMal]:EPSK:CONTrol <
Mode>, <Statistics
>
<Mode>
Description of parameters
Def. value
Def. unit
SCALar |
ARRay,
Scalar values only (incl. ramp matching)
Scalar measured values and arrays
ARRay –
<Statistics>
Description of parameters
Def. value
Def. unit
FW vers.
1 to 1000 |
OFF
Number of bursts per statistics cycle
Statistics off (equivalent to 1)
100 – V1.15
Description of command
This command restricts the type of measured values to accelerate the measurement and determines the number
of bursts within a statistics cycle.
CONFigure:POWer[:NORMal][:GMSK]:CONTrol:REPetition
Test
cycles
CONFigure:POWer[:NORMal]:EPSK:CONTrol:REPetition
<
Repetition
>
,<StopCond>,<Stepmode>
<Repetition>
Description of parameters
Def. value
Def. unit
CONTinuous |
SINGleshot |
1 to 10000
Continuous measurement (until
STOP
or
ABORT)
Single shot measurement (until
Status = RDY
)
Multiple measurement
(
counting
, until
Status = STEP | RDY)
SING –
<StopCondition>
Description of parameters
Def. value
Def. unit
SONerror |
NONE
Stop measurement in case of error
(stop on error)
Continue measurement even in case of error
NONE –
<Stepmode>
Description of parameters
Def. value
Def. unit
FW vers.
STEP |
NONE
Interrupt measurement after each statistics cycle
Continue measurement according to its rep. mode
NONE –
V1.15
Description of command
This command determines the number of statistics cycles, the stop condition and the stepping mode for the
measurement.
Note:
In the case of READ commands (
READ:
...), the <Repetition> parameter has no effect; the
measurement is always stopped after a single shot
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...