CMU-K20/K21/K22/K23/K24
Non Signalling: Modulation Measurements
1115.6088.12 4.53
E-11
Ovw., EVM, ME,
PE 8PSK
The
Ovw., EVM, ME, PE 8PSK
table section defines all limits for 8PSK-modulated
signals. The limits are set independently for the display modes
Current
and
Max./Min.
on one hand,
Average
on the other hand; see section
Measurement
Control (Modulation Configuration – Control)
on page
4.50
ff.
95
th
percentile PE
Upper limit for the phase error below which 95% of all
measured phase error values are located
95
th
percentile ME
Upper limit for the magnitude error below which 95% of all
measured relative magnitude error values are located
95
th
percentile EVM
Upper limit for the relative error vector magnitude below which
95% of all measured EVM values are located
Error Vector Magn.
Upper limits for the (peak and RMS-averaged
2
) relative error
vector magnitude (EVM). Both entries are positive.
Magnitude Error
Upper limits for the absolute value of the (peak and RMS)
relative magnitude error. Both entries are positive; the limits for
the peak magnitude error define a tolerance mask symmetric
to the origin.
Phase Error
Upper limits for the absolute value of the (peak and RMS)
phase error. Both entries are positive; the limits for the peak
phase error define a tolerance mask symmetric to the origin.
Origin Offset
Upper limit for the origin offset in the I/Q constellation diagram.
Frequency Error
Upper limit for the difference between the measured and the
expected frequency of the signal.
For an explanation of all measured quantities refer to the beginning of this section
on page
4.43
.
Remote control
CONFigure:MODulation:OEMP:EPSK:LIMit[:CURRent] ...
CONFigure:MODulation:OEMP:EPSK:LIMit:AVERage ...
Ext. Phase Error
GMSK
The table section
Ext. Phase Error GMSK
defines upper limits for the different
GMSK modulation parameters. The limits depend on the display mode of the
measurement curve:
Current & Max.
Common limits for the
Current
measurement curve and for the
Minimum/Maximum
curve (including the
Current
and the
Max./Min
scalar results)
Average
Limits for the
Average
measurement curve (including the
Average
scalar results)
For setting of the display mode see section
Measurement Control (Modulation
Configuration – Control)
on page
4.50
.
The meaning of the error limits is the same for the
Current
or
Minimum/Maximum
(Current & Max.)
and the
Average
results:
Phase Error Peak
Maximum phase error
Phase Error RMS
RMS phase error (RMS-averaged over the burst)
Origin Offset
Upper limit for the origin offset in the I/Q constellation diagram.
I/Q Imbalance
Upper limit for the amplitude difference between the in-phase
and quadrature components of the signal.
Frequency Error
Average frequency error in the burst
The
Phase Error Peak
and the
Frequency Error
are quantities with alternating sign;
the corresponding limits are symmetric to the origin (i.e. the absolute value of both
2
To keep the results comparable, RMS averaging was chosen for both positive quantities and quantities with alternating sign. The RMS-averaged EVM is calculated
according to the rule of GSM 05.05.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...