Connection Control (Non Signalling only) CMU-K20/K21/K22/K23/K24
1115.6088.12 6.4
E-12
CONFigure:RFANalyzer:MCONtrol:TSOFfset
<Slots>
Trig. Slot Offset
<Slots>
Description of parameters
Def. value
Def. unit
FW vers.
0 to 7
Trigger slot offset, no. of slots
0
–
V3.05
Description of command
This command defines a delay time between the trigger time and the measured timeslot.
RFANalyzer:DEFault <Enable>
Default Settings
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON |
OFF
The parameters are set to their default values
Some or all parameters differ from the default values
ON
– V3.05
Description of command
If used as a setting command with the parameter
ON
this command sets all parameters of the subsystem to their
default values (the setting
OFF
causes an error message).
If used as a query the command returns whether all parameters are set to their default values
(ON)
or not
(OFF)
.
Generator Object "RFGenerator" – Generator control
The subsystem
RFGenerator
configures the RF signals generated by the CMU. It corresponds to the
RF Generator
softkey in the
Generator
tab
in the popup menu
Connect. Control
.
INITiate:RFGenerator
Start RF generator, reserve resources
⇒
RUN
ABORt:RFGenerator
Switch off RF generator, release resources
⇒
OFF
Description of command
FW vers.
These commands have no query form. They start and stop the RF generator, setting it to the status
indicated in the top right column.
V1.15
FETCh:RFGenerator:STATus?
Generator
Returned values
Description of parameters
Def. value
Def. unit
FW vers.
OFF |
RUN |
ERR
Generator switched off (
ABORt
or
*RST)
Running (
INITiate
)
Switched off (could not be started)
OFF
– V1.15
Description of command
This command is always a query. It returns the current generator status.
Generator Level – Subsystem RFGenerator:LEVel
The subsystem
RFGenerator:LEVel
determines the level of the generated RF signals. It corresponds to
the input field
RF Level
of the panel
RF Generator
in the
Generator
tab
in the popup menu
Connect.
Control
.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...