CMU-K20/K21/K22/K23/K24
Measurement Groups (Non Signalling only)
1115.6088.12 6.21
E-12
Subsystem SUBarrays:POWer
The subsystem
SUBarrays:POWer
defines the measurement range and the type of output values.
CONFigure:SUBarrays:POWer[:NORMal][:GMSK]
Definition of Subarrays
CONFigure:SUBarrays:POWer[:NORMal]:EPSK
<Mode>,<Start>,<Samples>{,<Start>,<Samples>}
<Mode>
Description of parameters
Def. value
Def. unit
ALL |
ARIThmetical |
MINimum |
MAXimum |
IVAL,
Return all measurement values
Return arithm. mean value in every range
Return minimum value in every range
Return maximum value in every range
Return single interpolated value at <Start>
ALL –
<Start>
Description of parameters
Def. value
Def. unit
–10 bit to 156 ¾ bit,
Start time in current range (in bit for GMSK,
symbols for 8PSK modulation)
–10 bit
<Samples>
Description of parameters
Def. value
Def. unit
FW vers.
1 to 668
Number of samples in current range
668
–
V2.00
Description of command
This command configures the
READ:SUBarrays:POWer..., FETCh:SUBarrays:POWer...,
and
SAM-
Ple:SUBarrays:POWer
commands. It restricts the measurement to up to 32 subranges where either all meas-
urement results (the number of which is given by the second numerical parameter) or a single statistical value is
returned. The subranges are defined by the start time and the number of test points which are located on a fixed,
equidistant grid with a step width of ¼ bit. If <Start> does not coincide with a test point then the range will start at
the next test point that is larger than <Start>.
For <Mode> = IVAL, the <Samples> parameter is ignored and the CMU returns a single measurement value
corresponding to the abscissa value <Start>. If <Start> is located between two test points with valid results then
the result is calculated from the results at these two adjacent test points by linear interpolation.
The subranges may overlap but must be within the total range of the
POWer
measurement. Test points outside
this range are not measured (result
NAN)
and do not enter into the ARIThmetical, MINimum and MAXimum val-
ues.
By default, only one range corresponding to the total measurement range is used and all measurement values
are returned.
Measured Values – Subsystem POWer
The subsystem
POWer
determines and returns the results of the signal power measurement. They
correspond to the graphical measurement menu
Power
with its various display elements.
READ[:SCALar]:POWer[:NORMal][:GMSK]?
Scalar
results:
READ[:SCALar]:POWer[:NORMal]:EPSK?
Start single shot measurement and return results
FETCh[:SCALar]:POWer[:NORMal][:GMSK]?
FETCh[:SCALar]:POWer[:NORMal]:EPSK?
Read out measurement results (unsynchronized)
SAMPle[:SCALar]:POWer[:NORMal][:GMSK]?
SAMPle[:SCALar]:POWer[:NORMal]:EPSK?
Read out measurement results (synchronized)
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...