![Rohde & Schwarz CMU-K20 Operating Manual Download Page 461](http://html.mh-extra.com/html/rohde-and-schwarz/cmu-k20/cmu-k20_operating-manual_1477804461.webp)
CMU-K20/21/22/23/24 Signalling: Receiver Quality
1115.6088.12 6.157
E-12
Reference Level, Multislot
CONFigure:RXQuality:BER<nr>:CONTrol[:CSWitched][:TCH]:MSLot:RLEVel
<Level>
CONFigure:RXQuality:BER<nr>:CONTrol:PDATa:[:TCH]:MSLot:RLEVel
<Level>
<Level>
Description of parameters
Def. value
Def. unit
FW vers.
–
137 dBm to –27 dBm
–137 dBm to –10 dBm
–90 dBm to +13 dBm
Reference level for RF1
Reference level for RF2
Reference level for RF3 OUT
See below
See below
–90
dBm
dBm
dBm
V3.05
Description of command
This command defines the reference value for the individual downlink (BS) TCH signal levels used for the
multislot BER test on circuit switched channels. See command
CONFigure:RXQuality:BER<nr>:CONTrol[:CSWitched][:TCH]:MSLot:LEVel:INDividual
on p.
6.157
and the corresponding packet data command.
The default setting depends on the test setup (all level values in dB):
<nr>
1
2
3 4 5 6 7 8 9 10
Level –102.0
–104.0
–102.0 –104.0 –100.0 –100.0 –100.0 –102.0 –102.0 –102.0
Slot Configuration: Individual (Multislot)
CONFigure:RXQuality:BER<nr>:CONTrol[:CSWitched][:TCH]:MSLot:LEVel:INDividual
CONFigure:RXQuality:BER<nr>:CONTrol:PDATa[:TCH]:MSLot:LEVel:INDividual
<Level_0>, ..., <Level_7>
<Level_n>
Description of parameters
Def. value
Def. unit
FW vers.
–127 dB to +127 dB
Power of CMU in timeslot no. n
0
dB
V3.05
Description of command
This command defines the levels in all 8 timeslots of the downlink (BS) TCH signal relative to the
Reference
Level
set via
CONFigure:RXQuality:BER<nr>:CONTrol[:CSWitched][:TCH]:MSLot:RLEVel
(see p.
6.157
) and the corresponding packet data command. The levels are valid for BER tests if the MS is set to
multislot operation.
The level range quoted above is restricted by the condition that the absolute level (calculated from the reference
level and the relative individual levels) must not exceed the level ranges of the RF connectors.
Example:
With output connector RF2 and a default used timeslot level of –102 dBm, the individual levels can
be set in the range –35 dB to +92 dB, corresponding to absolute levels of –137 dBm to –10 dBm.
The PDATa command refers to packet-switched data traffic channels and requires option CMU-K42.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...