Common: POWer:XSLot
CMU-K20/21/22/23/24
1115.6088.12 6.46
E-12
Subsystem POWer:XSLot:CONTrol
CONFigure:POWer:XSLot:CONTrol:REPetition <
Repetition
>
,<StopCondition>,<Stepmode>
Test Cycles
<Repetition>
Description of parameters
Def. value
Def. unit
CONTinuous |
SINGleshot |
1 to 10000
Continuous measurement
(
continuous
, until
STOP
or
ABORT)
Single measurement
(single shot
, until
Status = RDY)
Multiple measurement
(
counting
, until
Status = STEP | RDY)
SING –
<StopCond>
Description of parameters
Def. value
Def. unit
SONerror |
NONE
Stop measurement in case of error
(stop on error)
Continue measurement even in case of error
NONE –
<Stepmode>
Description of parameters
Def. value
Def. unit
FW vers.
STEP |
NONE
Interrupt measurement after each statistics cycle
Continue measurement according to its rep. mode
NONE –
V2.15
Description of command
This command determines the number of statistics cycles and the stepping mode for the measurement.
Note:
In the case of READ commands (
READ:
...), the <Repetition> parameter has no effect; the
measurement is always stopped after a single shot.
CONFigure:POWer:XSLot:CONTrol:DEFault <Enable>
Default Settings
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON |
OFF
The parameters are set to their default values
The parameters differ from the default values (partially or
totally)
ON
– V2.15
Description of command
If used as a setting command with the parameter
ON
this command sets all parameters of the subsystem to their
default values (the setting
OFF
causes an error message).
If used as a query the command returns whether all parameters are set to their default values
(ON)
or not
(OFF)
.
Subsystem POWer:XSLot:SCOunt
The subsystem
POWer:XSLot:SOUnt
defines the total number of slots measured. It corresponds to the
Slot Count
parameter in the
Control
tab of the
Power Configuration
menu.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...