CMU-K20/21/22/23/24 Common:
POWer:SLOT
1115.6088.12 6.43
E-12
Subsystem SUBarrays:POWer:SLOT
The subsystem
SUBarrays:POWer:SLOT
defines the measurement range and the type of output
values.
CONFigure:SUBarrays:POWer:SLOT <Mode>,<Start>,<Samples>{,<Start>,<Samples>}
Definition of Subarrays
<Mode>
Description of parameters
Def. value
Def. unit
ALL |
ARIThmetical |
MINimum |
MAXimum |
IVAL,
Return all measurement values
Return arithm. mean value in every range
Return minimum value in every range
Return maximum value in every range
Return single interpolated value at <Start>
ALL –
<Start>
Description of parameters
Def. value
Def. unit
0 to 7,
Start time in current range
0
–
<Samples>
Description of parameters
Def. value
Def. unit
FW vers.
1 to 8
Number of samples in current range
8
–
V2.15
Description of command
This command configures the
READ:SUBarrays..., FETCh:SUBarrays...,,
and
SAMPle:SUBarrays:POWer:SLOT
commands. It restricts the measurement to up to 32 subranges where
either all measurement results (the number of which is given by the second numerical parameter) or a single
statistical value is returned. The subranges are defined by the current number of the first slot and the number of
slots within a subrange. If <Start> does not coincide with a test point then the range will start at the next test point
that is larger than <Start>.
For <Mode> = IVAL, the <Samples> parameter is ignored and the CMU returns a single measurement value
corresponding to the abscissa value <Start>. If <Start> is located beween two test points with valid results then
the result is calculated from the results at these two adjacent test points by linear interpolation.
The subranges may overlap but must be within the total range of the
P/Slot Graph
measurement. Test points
outside this range are not measured (result
NAN)
and do not enter into the ARIThmetical, MINimum and
MAXimum values.
By default, only one range corresponding to the total measurement range is used and all measurement values
are returned.
Subsystem POWer:SLOT
The subsystem
POWer:SLOT
measures power versus slot and returns the results. The subsystem
corresponds to the measurement menu
P/SLOT Graph.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...