CMU-K20/K21/K22/K23/K24
Non Signalling: Modulation Measurements
1115.6088.12 4.47
E-11
a) Scalar Results (Overview)
The measurement menu for the application
Overview 8PSK
shows all scalar results. Most of the values
are indicated in tabular form:
Parameter line
Output fields
Output table
and additional fields
Fig. 4-20 Display of results (Modulation – Overview)
Parameter line
The parameter line contains the following settings:
Max. Level
Maximum expected input level set as in the
Analyzer
tab of
the
Connection Control
menu
(see section
Table-Oriented
Version
on p.
4.73
ff.),
Attenuation
Setting for the external attenuation of the input level
(Normal,
Low Noise, Low Distortion)
,
Freq. Offset
Frequency offset compared to the nominal channel
frequency,
Chan./Trig. Slot Offs.
RF channel and trigger slot offset (see
Trig. Slot Offset
hotkey on p.
4.18
)
Remote control
The settings are retrieved using the query corresponding to the setting command
(setting command with appended question mark).
Output fields
In the output fields in the center of the menu, the following results are displayed:
TSC detected
Detected training sequence of the current burst received from
the mobile station
(GSM 0 to 7
or
Dummy
or
"---"),
see
Analyzer
tab, section
Table-Oriented Version
on p.
4.73
ff.
95
th
percentile
Limit values below which 95% of the measured
Error Vector
Magnitudes, Magnitude Errors,
and
Phase Errors
in the current
burst are located. Owing to this definition, the 95
th
percentile of a
measured quantity has the same unit as the quantity itself.
Output table
The scalar values in the output table are explained at the beginning of this section
on page
4.43
. They are first calculated for the current burst. From the current
results the average referenced to a statistics cycle (
Average,
see averaging rule in
Chapter 3, section
General Settings)
and the extreme value over all bursts
measured during the ongoing measurement
(Max/Min)
is calculated. Peak and
RMS
1
values are taken over the whole useful part of the burst.
1
To keep the results comparable, RMS averaging was chosen for both positive quantities and quantities with alternating sign. The RMS-averaged EVM is calculated
according to the rule of GSM 05.05.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...