CMU-K20/21/22/23/24
Options and Extensions
1115.6088.12 6.225
E-12
CONFigure:NETWork[:CSWitched]:AMR:HRATe:RSETting
AMR Rate Set, Half Rate
<CM4>, <CM3>, CM2>, <CM1>,
<ThrDown4>, <ThrUp3>, <ThrDown3>, <ThrUp2>,
<ThrDown2>, <ThrUp1>
<CM4>, <CM3>, <CM2>, CM1>
Description of parameters
Def. value
Def. unit
FW vers.
C1220 | C1020 | C0795 |
C0740 | C0670 | C0590 |
C0515 | C0475 |
OFF
User bit rate for codec modes 4 to 1.
Mode switched off
C0795,
C0670,
C0590,
C0515,
–
<ThrDown4>
Description of parameters
Def. value
Def. unit
0.0 dB to 31.5 dB
(in 0.5 dB steps)
Lower decision threshold for switching
between modes 4 and 3
15.0, dB
<ThrUp3>
Description of parameters
Def. value
Def. unit
0.0 dB to 31.5 dB
(in 0.5 dB steps)
Upper decision threshold for switching
between modes 3 and 4
17.0, dB
<ThrDown3>
Description of parameters
Def. value
Def. unit
0.0 dB to 31.5 dB
(in 0.5 dB steps)
Lower decision threshold for switching
between modes 3 and 2
12.5, dB
<ThrUp2>
Description of parameters
Def. value
Def. unit
0.0 dB to 31.5 dB
(in 0.5 dB steps)
Upper decision threshold for switching
between modes 2 and 3
15.0, dB
<ThrDown2>
Description of parameters
Def. value
Def. unit
0.0 dB to 31.5 dB
(in 0.5 dB steps)
Lower decision threshold for switching
between modes 2 and 1
11.0, dB
<ThrUp1>
Description of parameters
Def. value
Def. unit
FW vers.
0.0 dB to 31.5 dB
(in 0.5 dB steps)
Upper decision threshold for switching
between modes 1 and 2
13.0 dB V3.40
Description of command
Sig. State
These commands select four codec modes and define the decision thresholds for changing the codec
mode. The settings are valid for Half Rate AMR speech coder tests; see command
CONFigure:NETWork[:CSWitched]:SMODe:TRAFfic
on p.
6.197
. The instrument rejects the
settings (SCPI error –221, Settings conflict) unless the values meet all of the following conditions:
•
The rates must be in descending order so that <CM4> is the largest bit rate.
•
Up to 3 codec modes can be switched off. OFF must be the first values of the parameter list,
preceding the used codec modes.
•
Thresholds must be in descending order so that <ThrDown2>
≤
<ThrDown3>
≤
<ThrDown4> and
<ThrUp1>
≤
<ThrUp2>
≤
<ThrUp3>.
•
The hysteresis must be positive so that Up (j)
≥
Down (j + 1) for j= 1 to 3
To query the DL codec mode requested by the MS use
[SENSe:]MSSinfo:AMR:HRATe:DLCMode?
and
[SENSe:]MSSinfo:AMR:FRATe:DLCMode?
(see p.
6.209
).
≠
CEST,
Q: all
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...