Non Signalling: Power Measurements
CMU-K20/K21/K22/K23/K24
1115.6088.12 4.36
E-11
Remote control
CONFigure:POWer[:NORMal][:GMSK]:LIMit:LINE:UPPer:ALL:DYNamic:ENABle
CONFigure:POWer[:NORMal]:EPSK:LIMit:LINE:UPPer:ALL:DYNamic:ENABle
Static limit lines
The
Static
limit lines are defined as follows:
Area 1 to 16
Area
number
Enable
Enabled (switch on) or disabled (and invisible) limit line. A
disabled limit line implies that the limit check for the area is
switched off as well.
Time
Start and (below) stop time of the area in symbols
Level rel.
Start and (below) stop level of the area in units relative to the
carrier. The reference level (0-dB line) is the carrier power
averaged over the useful part of the burst.
Level abs.
Start and (below) stop level of the section in absolute units
(dBm).
The input of relative and absolute limit values is optional; both can be switched off
for valid areas (setting
Off)
. If both absolute and relative limit values are specified in
an area, the tolerance template and the results of the limit check refer to the
looser
criterion.
The permissible ranges for the upper and lower limit lines, i.e. of the quantities
Time, Level rel.,
and
Level abs
. vary according to the area numbers, see command
description in chapter 6.
Remote control
CONFigure:POWer[:NORMal][:GMSK]:LIMit:LINE:UPPer<nr>
[:STATic]:ENABle ON | OFF
CONFigure:POWer[:NORMal][:GMSK]:LIMit:LINE:UPPer<nr>[:STATic]
<StartTime>,<EndTime>,<StartRelLevel>,<EndRelLevel>,
<StartAbsLevel>,<EndAbsLevel>,<Visibility>
Dynamic limit
line correction
The
Dynamic
limit line section, serves to correct the limit lines depending on the
Template PCL
. It is defined as follows:
Range 1 to 10
Continuous range of power control levels defined by start PCL
and stop PCL
PCL from
Lowest template power control level in the range
PCL to
Highest template power control level in the range
Correction
Correction value in dB to be applied to the whole range
Enable
Enabled (switch on) or disabled dynamic correction
The dynamic limit line correction can be switched off entirely; see
Dynamic Limit
Line Correction
parameter above.
Remote control
CONFigure:POWer[:NORMal][:GMSK]:LIMit:LINE:UPPer<nr>:DYNamic:
ENABle ON | OFF
CONFigure:POWer[:NORMal][:GMSK]:LIMit:LINE:UPPer:ALL:DYNamic:
ENABle ON | OFF
CONFigure:POWer[:NORMal][:GMSK]:LIMit:LINE:UPPer<AreaNr>
:DYNamic<RangeNr>
<fromPCL>,<toPCL>,<Correction>,<Enable>
etc.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...