CMU-K20/K21/K22/K23/K24
Non Signalling: Connection Control
1115.6088.12 4.85
E-11
Fig. 4-39
Connection Control – Trigger
Default Settings
The
Default Settings
checkbox assigns the default setting to all functions in the
Trigger
tab (the default values are quoted in the command description in chapter 6
of this manual).
Remote control
TRIGger[:SEQuence]:DEFault ON | OFF
Meas. Trigger –
Source
Source
selects the source for the trigger event:
Free Run
Trigger by the GSM input signal: The CMU detects the burst; the
exact timing is given by the training sequence. This setting may
slow down the measurements.
P/t Multislot, P/Slot
, and
P/Frame
measurements can not be performed in
Free Run
trigger mode.
RF Power
The measurement is triggered by the level of the incoming burst
(rising or falling edge; see
Slope
setting below), the trigger level
is specified via the
Level
parameters. Wideband power trigger on
the RF Front End.
IF Power
The measurement is triggered by the level of the IF signal (rising
or falling edge; see
Slope
setting below), the trigger level is
specified via the
Level
parameters. Narrow-band IF power
trigger.
Extern
External trigger signal fed in via connector AUX 3 or AUX 4; see
Ext. Trigger ...
setting below.
For the
Free Run, RF Power
and
IF Power
settings the input signal must be a burst
signal. The external trigger can be selected for all
Non Signalling
measurements. In
contrast,
Signalling
measurements must be triggered by the signal from the
signalling unit or from the mobile phone.
RF Power
trigger signals have a small dynamic range which may not be sufficient
for triggering. It is recommended to trigger by the
IF Power
instead.
Note:
If no measurement result can be obtained the trigger mode may not fit
to the trigger signal applied. Check the trigger mode and signal.
Remote control
TRIGger[:SEQuence]:SOURce FRUN | EXTern
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...