CMU-K20/K21/K22/K23/K24
Non Signalling: Spectrum Measurements
1115.6088.12 4.57
E-11
The following parameter is specific to the
Spectrum due to Switching
measurement
and provided for both GSMK and 8PSK modulation:
Slot
Count
Slot Count
defines whether a single timeslot or all timeslots are considered for the
Spectrum due to Switching
measurement:
Single
The CMU measures the peak power in a fixed timeslot. The
measured timeslot is given by the trigger time plus the
Trig. Slot
Offset;
see
Fig. 4-3
on p.
4.18
.
A measurement cycle with
Statistic Count = n Bursts
extends over n TDMA frames, where
only the fixed timeslot, including the burst edges, is measured.
Multi
The CMU measures the peak power over all 8 timeslots of the
TDMA frame. The carrier output power (central bar in the
Spectrum due to Switching
diagram) is measured in the timeslot
defined by the trigger time plus the
Trig. Slot Offset
(Measured
Timeslot;
see
Fig. 4-3
on p.
4.18
)
.
A measurement cycle with
Statistic Count = n Frames
extends over n TDMA frames.
The single slot measurement is faster and is correct if the DUT operates in single
slot mode. The multislot measurement yields the correct
Spectrum due to
Switching
for any multislot configuration and for any levels in the individual UL
timeslots. The measured off-carrier power
does not depend on the
Measured
Timeslot,
however, the
Measured Timeslot
has an influence on the measured
carrier output power and thus on the limit lines (see
Table 4-3
on p.
4.65
). The
Measured Timeslot
can be changed in order to select the highest MS output power
as a reference for the tolerance template, in close analogy to the single slot mode.
Remote control
CONFigure:SPECtrum:SWITching[:GMSK]:SCOunt SSL | MSL
CONFigure:SPECtrum:SWITching:EPSK:SCOunt SSL | MSL
Appli-
cation
The
Application
softkey changes the type of spectrum to be measured and the
modulation scheme. The two alternative spectra (applications) are displayed in
separate measurement menus. When an application is selected, the corresponding
measurement menu is called up and the labeling of the measurement control
softkey is adapted. The configuration settings for both applications, however, are
listed in a common popup menu (see p.
4.61
ff.).
Modulation
GMSK
The
Modulation GMSK
hotkey selects the spectrum due to modulation for GMSK
modulated signals to be displayed. In this mode, the power at the nominal
frequency and the relative power at up to 11 neighboring frequencies is displayed
in a single graph, see p.
4.58
ff.
Remote control
No explicit switchover command. All spectrum due to modulation measurements
are identified by the 2
nd
to 4
th
level keywords
..SPECtrum:MODulation[:GMSK]
Modulation
8PSK
The
Modulation 8PSK
hotkey selects the spectrum due to modulation for 8PSK
modulated signals to be displayed. In this mode, the power at the nominal
frequency and the relative power at up to 11 neighboring frequencies is displayed
in a single graph, see p.
4.58
ff.
Remote control
No explicit switchover command. All spectrum due to modulation measurements
are identified by the 2
nd
to 4
th
level keywords
..SPECtrum:MODulation:EPSK
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...