Signalling: Receiver Quality
CMU-K20/21/22/23/24
1115.6088.12 6.166
E-12
CONFigure:RXQuality:BAVerage:CONTrol:DEFault <Enable>
Default Settings
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON |
OFF
The parameters are set to default values
Some or all parameters differ from the default value
ON
– V2.00
Description of command
As a
setting command
with the setting
ON
this command sets all parameters of the subsystem to default values
(the setting
OFF
causes an error message).
As a query, this command reads out whether all parameters are set to default values
(ON)
or not
(OFF)
.
Subsystem RXQuality:BAVerage:LIMit
The subsystem
RXQuality:BAVerage:LIMit
defines tolerance values for the continuous receiver quality
measurement. The subsystem corresponds to the tab
Limits
in the popup menu
Receiver Quality
Configuration.
CONFigure:RXQuality:BAVerage:LIMit:CLII
<ClassIIBER>
Class II Bits
<ClassIIBER>
Description of parameters
Def. value
Def. unit
FW vers.
0 % to 100 %
Upper limit of error rate for class II bits
0.2
%
V2.00
Description of command
This command defines an upper limit for the bit error rate of class II (unprotected bits, see chapter 4) for
continuous application.
CONFigure:RXQuality:BAVerageLIMit:CLIB
<ClassIbBER>
Class Ib Bits
<ClassIbBER>
Description of parameters
Def. value
Def. unit
FW vers.
0 % to 100 %
Upper limit of error rate for class Ib bits
0.4
%
V2.00
Description of command
This command defines an upper limit for the bit error rate of class Ib (partly protected bits, see chapter 4) for the
continuous application.
CONFigure:RXQuality:BAVerage:LIMit:FERRors <Frame Errors>
Frame Errors
<FERRors>
Description of parameters
Def. value
Def. unit
FW vers.
0 % to 100 %
Upper limit for erased frame errors
0.1
%
V2.00
Description of command
This command defines an upper limit for the relative portion of invalid and therefore erased frames (
frame
erasure rate
, see chapter 4) in the measurement of the residual bit error rate (
RBER
, see command
CONFigure:RXQuality:BAVerage:CONTrol
) and for the continuous application.
CONFigure:RXQuality:BAVerage:LIMit:DBLer <Data_BLER>
Data Block Error Rate
<Data_BLER>
Description of parameters
Def. value
Def. unit
FW vers.
0 % to 100 %
Upper limit for data block error rate
10.0
%
V3.05
Description of command
This command defines an upper limit for the data BLER.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...