GPRS Signalling and EGPRS
CMU-K20/K21/K22/K23/K24
1115.6088.12 4.196
E-12
The CMU provides a softkey-oriented version of the
MS Signal
tab and a table-oriented version with
extended functionality. The
MS Signal
hotkey toggles between the two versions if it is pressed
repeatedly. For packet data mode, the two tabs provide the same settings.
Fig. 4-88 Connection Control – MS Signal (table)
PMAX
The
PMAX
parameter sets the maximum MS transmitter output power allowed in
the cell. The value corresponds to the output power at which the mobile station
performs a location update to synchronize to the CMU. It is valid both for circuit
switched and for packet data mode.
Remote control
CONFigure:MSSignal:CCH:PMAX_
Packet Data –
Multi Slot
The
Packet Data – Multi Slot
section defines the timeslot configuration in the uplink
and downlink:
Main Timeslot
Timeslot used for signalling. The main timeslot can not be
switched off in both the downlink and uplink; see Slot
Configuration below. In Receiver Quality tests the main timeslot
is always one of the measured slots.
PCL (MS)
PCL (MS) sets (signalling states < Call Established) or changes
(signalling states Call Established) the MS output power during the connection.
where is this setting?
Remote control
CONFigure:BSSignal:PDATa[:TCH]:MSLot:MTIMeslot
Slot
Configuration
Table of all used and unused timeslots (GSM timeslots 0 to 7) in the downlink and
the uplink. The
Main Timeslot
is always active in both the downlink and uplink
direction. The boxes enable (if checked) or disable the other timeslots.
Level (BS)
RF levels in the individual downlink timeslots (RF signal
transmitted by the BS/CMU) relative to the
Reference Level
indicated above the
Slot Configuration
table and in the
BS Signal
tab (see section see section
RF Signals of the CMU (Connection
Control – BS Signal)
on p.
4.197
f.).
Gamma
Channel-specific power control parameter
Γ
CH
in dB; see note on
GPRS power control above
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...