Common: MODulation:PERRor
CMU-K20/21/22/23/24
1115.6088.12 6.92
E-12
Test Configuration
The commands of the following subsystems configure the
Modulation
measurement. They correspond
to the sections in the
Modulation Configuration
menu that are related to the
Phase Error
application.
Subsystem MODulation[:PERRor]:EPSK:CONTrol
The subsystem
MODulation[:PERRor]:EPSK:CONTrol
configures the modulation measurement. It cor-
responds to the
Control
tab in the popup menu
Modulation Configuration
.
CONFigure:MODulation[:PERRor]:EPSK:CONTrol <
Mode>, <Statistics
>
Scope of Measurement
<Mode>
Description of parameters
Def. value
Def. unit
SCALar |
ARRay
Only scalar measured values (incl. tolerance matching)
Scalar measured values and arrays
ARRay –
<Statistics>
Description of parameters
Def. value
Def. unit
FW vers.
1 to 1000 |
NONE
No. of bursts within a statistics cycle
Statistics off
100 – V2.15
Description of command
This command selects the type of measured values and determines the number of bursts forming one statistics
cycle.
CONFigure:MODulation[:PERRor]:EPSK:CONTrol:REPetition
<
Repetition
>
,<StopCond>,<Stepmode>
Test Cycles
<Repetition>
Description of parameters
Def. value
Def. unit
CONTinuous |
SINGleshot |
1 to 10000
Continuous measurement (until
STOP
or
ABORT)
Single shot measurement (until
Status = RDY
)
Multiple measurement
(
counting
, until
Status = STEP | RDY)
SING –
<StopCond>
Description of parameters
Def. value
Def. unit
SONerror |
NONE
Stop measurement in case of error
(stop on error)
Continue measurement even in case of error
NONE –
<Stepmode>
Description of parameters
Def. value
Def. unit
FW vers.
STEP |
NONE
Interrupt measurement after each statistics cycle
Continue measurement according to its rep. mode
NONE –
V2.15
Description of command
This command determines the number of statistics cycles, the stop condition and the stepping mode for the
measurement.
Note:
In the case of READ commands (
READ:
...), the <Repetition> parameter has no effect; the measure-
ment is always stopped after a single shot.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...