Signalling: Receiver Quality
CMU-K20/21/22/23/24
1115.6088.12 6.152
E-12
Receiver Quality
The subsystem
RXQuality
comprises the commands for all receiver quality measurements. The
subsystem corresponds to the main menu
Receiver Quality
and the associated popup menu
Receiver
Quality Configuration.
Important Note: Receiver Quality Measurements with MCS-5 to MCS-9
Receiver Quality measurements on circuit-switched channels using the modulation and coding
schemes MCS-5 to MCS-9 (see
CONFigure:NETWork[:CSWitched]:SMODe:TRAFfic
command) can not be performed in parallel to TX Tests
(POWer...,
MODulation...,SPECtrum)
. The Receiver Quality measurement must be in the READY or
OFF state before a TX Test can be started. Conversely, all TX Tests must be in the READY or
OFF state before a Receiver Quality measurement using MCS-5 to MCS-9 can be started.
General Settings – Subsystem RXQuality...
The subsystem
RXQuality...
provides settings for all types of receiver quality measurements.
CONFigure:RXQuality[:CSWitched]:BITStream <
Mode
>
Bit Stream BER
CONFigure:RXQuality:PDATa:BITStream <
Mode
>
<Mode>
Description of parameters
Def. value
Def. unit
FW vers.
PR9 |
PR11 |
PR15 |
PR16
2
9
-1 PSR bit pattern
2
11
-1 PSR bit pattern
2
15
-1 PSR bit pattern
2
16
-1 PSR bit pattern
PR9 –
V3.10
Description of command
Sig. State
This command defines the pseudo random bit sequence that the CMU transmits to the MS during
Receiver Quality
measurements.
all
CONFigure:RXQuality:CONTrol:HTIMe <AGCTime>, <
SynchTime>
AGC Holdoff Time, Sync Holdoff Time
<AGCTime>
Description of parameters
Def. value
Def. Unit
0 s to 100 s
Automatic gain control
0.5
s
<SynchTime>
Description of parameters
Def. value
Def. Unit
FW vers.
0 s to 100 s
Bit stream synchronization
0.2
s
V2.00
Description of command
This command defines hold off times during which the mobile can adapt itself to the new RF level at the
beginning of the
Receiver Quality
measurement and send back the bit stream received.
By reducing the hold off times, the measurement can be speeded up. The default settings for both times do not
depend on the application.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...