Common: SPECtrum:MODulation
CMU-K20/21/22/23/24
1115.6088.12 6.104
E-12
FETCh:SPECtrum:MODulation[:GMSK]:STATus?
FETCh:SPECtrum:MODulation:EPSK:STATus?
Measurement Status
Returned values
Description of parameters
Def. value
Def. unit
FW vers.
OFF |
RUN |
STOP |
ERR |
STEP |
RDY,
1 to 10000 |
NONE,
1 to 1000 |
NONE
Measurement in the
OFF
state (
*RST
or
ABORt
)
Running (after
INITiate, CONTinue
or
READ
)
Stopped (
STOP
)
OFF
(could not be started)
Stepping mode
(<stepmode>=STEP)
Stopped according to repetition mode and stop condition
Counter for current statistics cycle
No counting mode set
Counter for current evaluation period within a cycle
Statistic count set to off
OFF
NONE
NONE
–
–
–
V1.20
Description of command
This command is always a query. It returns the status of the measurement (see chapters 3 and 5 of CMU man-
ual).
Subsystem SPECTrum:MODulation:CONTrol
The subsystem
SPECtrum:MODulation:CONTrol
defines the repetition mode, statistic count, and stop
condition of the measurement. These settings are provided in the
Control
and
Meas X
tabs of the popup
menu
Spectrum Configuration
.
CONFigure:SPECtrum:MODulation[:GMSK]:CONTrol <
Mode>,<Statistics
>
CONFigure:SPECtrum:MODulation:EPSK:CONTrol <
Mode>,<Statistics
>
Scope of Measurement
<Mode>
Description of parameters
Def. value
Def. unit
SCALar |
ARRay
Only scalar measured values
Scalar measured values and arrays
ARRay –
<Statistics>
Description of parameters
Def. value
Def. unit
FW vers.
1 to 1000 |
NONE
Number of bursts per statistics cycle
Statistics off (equivalent to 1)
200 – V1.20
Description of command
This command restricts the type of measured values and determines the number of bursts within a statistics cy-
cle.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...