Non Signalling: Power Measurements
CMU-K20/K21/K22/K23/K24
1115.6088.12 4.8
E-11
Power Measurements
The menu group
Power
is designed to measure the RF output power of the MS transmitter. The power
can be analyzed as a function of time in a single timeslot or in up to 4 consecutive timeslots.
Furthermore, the CMU evaluates the average power and its evolution over several consecutive slots or
frames. The different measurements are treated as different applications; the results are displayed in
separate
Power
measurement menus.
The popup menu
Power Configuration
is used to configure the
measurements.
The different types of Power measurements are treated as applications which can be selected with the
Application
softkey.
P/t Normal
The
P/t Normal ...
(burst power versus time) application measures the output power of the
DUT over one burst period. The measurement curve obtained can be further processed to
determine an average, minimum, or maximum result and calculate the average over the
whole burst.
P/t
measurements are provided for normal bursts at GMSK or 8MSK
modulation.
In addition to the burst power measurement, a limit check with tolerances depending on the
RF output power of the DUT and the modulation scheme is performed; see section
Limit
lines (Power Configuration – Limit Lines)
on p.
4.32
ff.
P/t Multislot
The
P/t Multislot
application measures the output power of the DUT over up to 653 symbol
periods, corresponding to 4 timeslots plus an appropriate display margin. This
measurement is particularly suited to GSM multislot solutions like GPRS or circuit-switched
HSCSD where several timeslots can be allocated to a single connection.
The multislot measurement curve can be further processed to determine an average,
minimum, or maximum result and calculate the average power over each burst measured.
P/t Multislot
measurements are provided both in
Non Signalling
and in
Signalling
test mode
and for normal bursts at GMSK and 8PSK modulation. In addition to the burst power
measurement, a limit check with tolerances depending on the RF output power of the DUT
and the modulation scheme is performed; see section
Limit lines (Power Configuration
– Limit Lines)
on p.
4.32
ff.
P/Slot
The
P/Slot
applications measure the average burst power in a series of consecutive
timeslots. The average is taken over a section of the useful part of the burst; it is not
correlated to the training sequence. The result is displayed either in a bar graph (all eight
timeslots of a single TDMA frame,
P/Slot Graph
application) or in a table (up to 512
timeslots, corresponding to a total test time of approx. 0.3 s,
P/Slot Table
application).
P/Frame
The
P/Frame
measurement represents a fast and convenient method of monitoring the
behavior of the average burst power in a particular timeslot over a whole range of
consecutive TDMA frames. The measurement extends over a range of up to 256 frames,
corresponding to test times of less than 1.2 s. The average is obtained like in the case of
P/Slot
measurements.
The
P/Slot
and the
P/Frame
measurement complement the
P/t
measurement where a large
number of bursts can be measured but the output of the average burst power is restricted to
current, average, minimum or maximum values within a statistics cycle (see
Display Mode
setting below).
P/Slot and P/Frame
returns
all
values; the applications are suitable
whenever the behavior or the stability of the average burst power in particular timeslots are
to be monitored over an extended time range in R&D.
RF Level
In all applications, the CMU measures at arbitrary RF input levels provided that they are
within the allowed range of the RF input connectors.
Signalling mode
Note:
In Signalling mode, where the CMU is able to test a broad range of signalling
issues, two further measurement applications are available (see section
Power Measurements
on page
4.97
ff):
The average burst power can be measured as a function of the PCL of the
mobile phone (application
P/PCL)
.
Access bursts from the mobile station can be measured (see also section
Limit lines (Power Configuration – Limit Lines)
on page
4.32
ff.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...