Non Signalling: Power Measurements
CMU-K20/K21/K22/K23/K24
1115.6088.12 4.10
E-11
a) Measurement Control
Each
Power
application is controlled by means of the measurement control softkey below the
Connect.
Control
softkey and the associated hotkeys.
P/t Norm.
GMSK
The
P/t Norm. GMSK
measurement control softkey (which changes to
P/t Norm.
8PSK
etc., depending on the application selected) controls the power
measurement application and indicates its status
(RUN | HLT | OFF)
. This status
can be changed after softkey selection (pressing once) by means of the
ON/OFF
key or the
CONT/HALT
key. The status can be set independently for all
Power
applications.
The active
Power
application generally suspends the other applications. On
switchover between different applications, the selected measurement status of
each application is stored and will be put into effect as soon as the application is
activated. In particular, an application in the status
RUN
is restarted each time it is
activated.
Remote control
INITiate:<Application>
ABORt:<Application>
STOP:<Application>
CONTinue:<Application>
FETCh:<Application>:STATus?
<Application> = POWer:NORMal[:GMSK]
etc.
Measurement
configuration
Pressing the
P/t Norm. GMSK
softkey twice opens the popup menu
Power
Configuration
(see page
4.28
). Besides, the measurement control
softkey provides
hotkeys to define the scope of the measurement. All these settings are described in
more detail in section
Measurement Control (Power Configuration – Control)
on
page
4.28
ff.
b) Selecting the Measurement Application
Appli-
cation
The Application softkey selects the power measurement application.
The applications
P/t Normal
<Mod_Type>
depend on the modulation scheme of the
analyzed signal. In the
P/t Multislot
application, the modulation in each measured
slot can be defined separately. The
P/Frame, P/Slot Graph,
and
P/Slot Table
menus and settings do not depend on the modulation scheme.
The
Power
measurement menu and the measurement control softkey change with
the application selected; the results are explained in section
Measurement Results
on page
4.20
ff.
P/t Normal
GMSK
The
P/t Normal GMSK
hotkey selects the power versus time measurement for
GMSK modulated normal burst signals (see explanation of GSM burst structure at
the beginning of section
Limit lines (Power Configuration – Limit Lines)
on page
4.32
).
Remote control:
The
P/t Normal GMSK
application is selected by the keywords [:NORMal] [:GMSK]
in the 3
rd
and 4
th
level of the POWer commands, e.g. CONFigure:POWer
[:NORMal][:GMSK]...
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...