Non Signalling: Power Measurements
CMU-K20/K21/K22/K23/K24
1115.6088.12 4.30
E-11
Continuous
Display of minimum and maximum from all bursts already
measured. The
average value
, however, is calculated according
to the rule in Chapter 3, section
General Settings
.
Remote control
no display mode set explicitly, the four measurement curves are accessible via
FETCh:ARRAy:POWer[:NORMal][:GMSK][:CURRent]?
FETCh:ARRAy:POWer[:NORMal][:GMSK]:MINimum?
FETCh:ARRAy:POWer[:NORMal][:GMSK]:MAXimum?
FETCh:ARRAy:POWer[:NORMal][:GMSK]:AVERage?
etc.
Statistic Count
The input field
Statistic Count
defines the length of the statistics cycles in bursts.
The settings
1
and
OFF
(press
ON/OFF
key) are equivalent. A statistics cycle
determines the duration of single-shot measurements.
Remote control
CONFigure:POWer[:NORMal][:GMSK]:CONTrol
<MODE>,1 ... 1000 | NONE
Filter
The input fields
Filter
determine which type of measurement filter is used for the P/t
measurements:
500 kHz Gauss
Gauss filter with a 3-dB bandwidth of 500 kHz, recommended for
GMSK modulation
600 kHz Band
Bandpass filter with a bandwidth of 600 kHz and steep edges,
recommended for 8PSK modulation
Both filters are in accordance with the conformance specification GSM 11.10.
Remote control
CONFigure:POWer[:NORMal][:GMSK]:FILTer G500 | B600
CONFigure:POWer[:NORMal]:EPSK:FILTer G500 | B600
CONFigure:POWer:MSLot:FILTer G500 | B600
Grid
The
Grid
button switches the grid on or off in the graphical test diagram. By default,
the grid is switched on.
Remote control
CONFigure:POWer[:NORMal][:GMSK]:CONTrol:GRID ON | OFF
The following settings are application-specific:
P/t Normal
8PSK, P/t 8PSK,
Multislot –
Ref. Power Mode
The
Ref. Power Mode
determines how the reference power, i.e. the 0-dB line in the
measurement diagram, is calculated. The setting is valid for
both
single-slot and
multislot measurements on 8PSK modulated signals.
Current
The reference power depends on the
Display Mode
set. It is
equal to the average power of the
Current
measurement curve
(display mode
Current)
or to the average power of the
Average
measurement curve (display mode
Average, Maximum, or
Minimum)
.
Average
The reference power is equal to the average power of the
average measurement curve.
Data Compens.
The reference power depends on the
Display Mode
set. It is
equal to the data-compensated average power of the
Current
measurement curve (display mode
Current)
or to the data-
compensated average power of the
Average
measurement curve
(display mode
Average, Maximum, or Minimum)
.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...