CMU-K20/K21/K22/K23/K24
Non Signalling: Analyzer/Generator Menu
1115.6088.12 4.5
E-11
Selecting the Application
Appli-
cation
The
Application
softkey selects the measurement application. The measurement
control softkey (second softkey below
Connect. Control)
indicates the current
application. Some of the hotkeys associated to the different softkeys, the
Setup
table, and the results in the
Analyzer/Generator
menu also vary as a function of the
application. The corresponding measurement results are explained in section
Measurement Results
on page
4.6
ff.
P/t Normal
GMSK
The
P/t Normal GMSK
hotkey selects the power versus time measurement for
normal burst signals. See section
Power Measurements
on p.
4.8
.
Remote control
The
P/t Normal GMSK
application is selected by the keywords
[:NORMal] [:GMSK]
in the 3
rd
and 4
th
level of the
POWer
commands, e.g.
CONFigure:POWer [:NORMal][:GMSK]...
Ext. Phase
Err. GMSK
The
Ext. Phase Error GMSK
hotkey selects the measurement of the modulation
accuracy of GMSK modulated signals. See section
Measurement Menu
(Modulation – GMSK)
on p.
4.38
.
Remote control
The
Phase Error GMSK
application is selected by the keywords
XPERror[:GMSK]
in the 3
rd
and 4
th
level of the
MODulation
commands, e.g.
CONFigure:MODulation:XPERror[:GMSK]...
Overview
8PSK
The
Overview 8PSK
hotkey selects the measurement of the power and modulation
accuracy of 8PSK modulated signals. See section
Measurement Menu (Modulation
– 8PSK)
on page
4.42
.
Remote control
The
Overview 8PSK
application is selected by the keywords
OVERview:EPSK
in
the 3
rd
and 4
th
level of the
MODulation
commands, e.g.
CONFigure:MODulation:OVERview:EPSK...
Application-Specific Settings
As outlined in section
Analyzer/Generator Menu
on p.
4.2
, some of the hotkey/softkey combinations in
the
Analyzer/Generator
menu vary as a function of the application. However, all
Analyzer/Generator
settings are always identical to the corresponding settings in the
Power
and
Modulation
menus.
Changes made in the
Analyzer/Generator
menu overwrite the
Power
and
Modulation
settings and vice
versa.
Description
of settings
•
The settings to be made in the
P/t Normal GMSK
application are described in
section
P/t Normal GMSK
on p.
4.11
ff.
•
The settings to be made in the
Ext. Phase Error GMSK
application are
described in section
Test Settings
on p.
4.39
ff.
•
The settings to be made in the
Overview 8PSK
application are described in
section
Test Settings
on p.
4.45
ff.
Setup table
The
Setup
table in the right half of the
Analyzer/Generator
menu gives an overview
of the measurement settings belonging to the current application. It changes when
a different application is selected. The roll-key scrolls and expands the
Setup
table.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...