CMU-K20/K21/K22/K23/K24
Non Signalling: Modulation Measurements
1115.6088.12 4.41
E-11
Measurement Results
The values shown in the measurement menu
Modulation
can be divided into three groups:
•
Setting
values
•
Scalar measurement results (single values)
•
Arrays (the measurement curve plotted as a function of time)
The values are indicated in a parameter line, the test diagram and a tabular overview below:
Parameter line 1,2
Test diagram
Output fields and
table
Fig. 4-16 Display of results (Modulation – Ext. Phase Err. GMSK)
Settings/
Scalar results
Scalar measurement results and settings are indicated in the two parameter lines
above the test diagram and in the output table below.
1
st
parameter line
The first parameter line contains the following settings:
Max. Level
Maximum input level set as in the
Analyzer
tab of the
Connection Control
menu
(see section
Table-Oriented
Version
on p.
4.73
ff.)
Attenuation
Setting for the external attenuation of the input level
(Normal,
Low Noise, Low Distortion)
Freq. Offset
Frequency offset compared to the nominal channel frequency
Chan./Trig. Slot Offs.
RF channel and trigger slot offset (see
Trig. Slot Offset
hotkey on p.
4.18
)
2
nd
parameter
line
The second parameter line contains the following marker values:
Level and time of reference marker
Level and time of delta marker 1 (setting
absolute
) or difference
from reference marker (setting
relative
)
Level and time of delta marker 2 (setting
absolute
) or difference
from reference marker (setting
relative
)
Remote control
The settings are retrieved using the query corresponding to the setting command
(setting command with appended question mark).
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...