![Rohde & Schwarz CMU-K20 Operating Manual Download Page 336](http://html.mh-extra.com/html/rohde-and-schwarz/cmu-k20/cmu-k20_operating-manual_1477804336.webp)
Common: Option Query
CMU-K20/21/22/23/24
1115.6088.12 6.32
E-12
Common Measurements and Command Groups
The commands for the measurement groups in this section are identical or almost identical in both test
modes. Minor differences between
Non Signalling
and
Signalling
commands are possible; they will be
pointed out throughout the section.
Note1: Measurements
and signalling states
To perform any kind of measurement and obtain a meaningful result, an appropriate test
setup is required (see application examples in chapter 2 of this manual). Consequently, if
the measurements reported in this section are performed in Signalling mode, the Call Es-
tablished (CEST) signalling state must be reached before any of the commands retrieving
test results
(READ...?, FETCh...?
,
SAMPle...?
, or
CALCulate...LIMit?)
can be
used. Test configurations, however, can be defined any time.
Exception: The wideband power
(WPOWer)
does not involve any signalling. It can be
measured irrespective of the current signalling state.
Note2:
GMSK and 8PSK modulation
The
keywords
[:GMSK]
and
:EPSK
in the remote control commands denote GMSK and
8PSK modulation, respectively. The
:EPSK
commands in Signalling measurements are in-
cluded in firmware versions V3.05 and higher. The firmware version numbers quoted in the
command tables refer either to GMSK modulation or EPSK modulation in Non Signalling
mode.
Option Query
The
Options
subsystem contains the commands for querying information on the instrument and the
available options. It corresponds to the
Options
tab in the
Setup
menu opened via the
SETUP
key on
the front panel.
SYSTem:OPTions:INFO:CURRent?
Device Info
Response
Def. value
Default unit
FW vers.
Example:
Rohde&Schwarz,CMU 200-1100.0008.02,840675/018,
V3.10C:SP02
2002-09-05"GSM900MS_Sig"
–
– V3.10
Command description
This command returns the information on the device comprising the manufacturer, model, serial number and
firmware version of the current function group. This command is always a query.
深圳德标仪器
135-1095-0799
Summary of Contents for CMU-K20
Page 4: ...深圳德标仪器 135 1095 0799 ...
Page 8: ...深圳德标仪器 135 1095 0799 ...
Page 10: ...深圳德标仪器 135 1095 0799 ...
Page 12: ...深圳德标仪器 135 1095 0799 ...
Page 18: ...深圳德标仪器 135 1095 0799 ...
Page 24: ...深圳德标仪器 135 1095 0799 ...
Page 26: ...深圳德标仪器 135 1095 0799 ...
Page 34: ...深圳德标仪器 135 1095 0799 ...
Page 58: ...深圳德标仪器 135 1095 0799 ...
Page 60: ...深圳德标仪器 135 1095 0799 ...
Page 78: ...深圳德标仪器 135 1095 0799 ...
Page 286: ...深圳德标仪器 135 1095 0799 ...
Page 288: ...深圳德标仪器 135 1095 0799 ...
Page 304: ...深圳德标仪器 135 1095 0799 ...
Page 632: ...深圳德标仪器 135 1095 0799 ...